Energy dependence of TLD-300 response from 6 keV up to 1250 keV

被引:3
作者
Angelone, M
Chiti, M
Esposito, A
Fantuzzi, E
Monteventi, F
机构
[1] EURATOM ENEA, Ctr Ric Frascati, I-00044 Frascati, Italy
[2] Ist Nazl Fis Nucl, Lab Nazl Frascati, I-00044 Frascati, Italy
[3] ENEA, Ist Radioprot, I-40136 Bologna, Italy
关键词
D O I
10.1093/oxfordjournals.rpd.a005894
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The energy dependence of the response function S(E), of CaF2:Tm (TLD-300) thermoluminescent material (3.2 X 3.2 X 0.90 mm(3) ribbons) was measured from 6 keV LIP to 1250 keV (Co-60) using Fe-55, Cs-137 and Co-60 gamma ray sources as well as X ray ISO reference beams. To the best knowledge of the authors the data for energies below 45 keV are reported for the first time. The S(E) function was also calculated in the energy range from I keV up to 20 MeV and compared with the experimental data relative to the total glow curve. The comparison shows good agreement front 1250 keV down to 40 keV. At lower energy the discrepancy increases. This is interpreted in terms of the X ray attenuation within the thickness of TLD-300. Finally, measurements point out the effects of different reading and annealing treatments on the measured function S(E).
引用
收藏
页码:381 / 384
页数:4
相关论文
共 11 条
[1]   SENSITIVITY OF LIF THERMOLUMINESCENT DOSIMETERS TO 6-18 KEV PHOTONS [J].
AHMED, AB ;
BARBER, DE .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (03) :343-352
[2]   ENERGY RESPONSE OF SOME THERMOLUMINESCENT DOSIMETERS TO 4-12 KEV X-RAYS [J].
BACCI, C ;
CANNATA, A ;
ESPOSITO, A ;
FURETTA, C ;
PELLICCIONI, M .
RADIATION EFFECTS LETTERS, 1981, 67 (1-2) :19-25
[3]   CALCULATED ENERGY-DEPENDENCE OF SOME RTL AND RPL DETECTORS [J].
BASSI, P ;
BUSUOLI, G ;
RIMONDI, O .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1976, 27 (5-6) :291-305
[4]  
BOS AJJ, 1991, RADIAT PROT DOSIM, V37, P231
[5]   Dose response of the individual peak for LiF:Mg,Ti (TLD-100) and LiF:Mg,Cu,P (GR-200A) TL materials at the photon energies of 30, 104, 1250 keV [J].
Cai, GG ;
Bestion, N ;
Wang, SS ;
Shen, WX ;
Delard, R ;
Volpiliere, J ;
Fesquet, J ;
Gasiot, J .
RADIATION PROTECTION DOSIMETRY, 1996, 65 (1-4) :213-215
[6]  
FURETTA C, 1985, RADIAT PROT DOSIM, V11, P101
[7]  
Hubbel J H, 1995, 5632 NISTIR
[8]  
*INT ORG STAND, 1997, X GAMM REF RAD CAL D
[9]  
McKeever S. W. S., 1995, Thermoluminescence Dosimetry Materials: Properties and Uses
[10]   SELF-ABSORPTION EFFECT IN LIF THERMOLUMINESCENT DOSIMETERS [J].
MUKHERJEE, B ;
VANA, N .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 226 (2-3) :572-573