Concentration, size, mean lifetime, and noise effects on image quality in luminescence optical tomography

被引:1
作者
Chang, J
Graber, HL
Barbour, RL
机构
来源
OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE: THEORY, INSTRUMENTATION, MODEL, AND HUMAN STUDIES II, PROCEEDINGS OF | 1997年 / 2979卷
关键词
D O I
10.1117/12.280221
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The impact of background lumiphore in luminescence optical tomography is examined. To demonstrate its effects, numerical simulations were performed to calculate the diffusion-regime limiting form of forward-problem solutions for a specific test medium. Image reconstructions were performed using a CGD algorithm with a rescaling technique and positivity constraints. In addition, we develop a modification to the basic algorithm that makes use of the maximum possible concentration in order to estimate the background concentration, and show that it improves Image quality when background luniphore is present. We conclude that the usual measure of background lumiphore's effect which is the target-to-background lumiphore concentration ratio, is not adequate to define the contribution from the background lumiphore. The reason for this is that image quality is also a function of target-size and location. An alternative measure that we find superior is described. The results indicate that the improved algorithm yields better image quality for low target-to-background ratios.
引用
收藏
页码:750 / 758
页数:9