Design, fabrication, and microfabricated columns evaluation of for gas chromatography

被引:139
作者
Lambertus, G
Elstro, A
Sensenig, K
Potkay, J
Agah, M
Scheuering, S
Wise, K
Dorman, F
Sacks, R [1 ]
机构
[1] Univ Michigan, Dept Chem, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Elect Engn & Comp Sci, Ann Arbor, MI 48109 USA
[3] Muhlenberg Coll, Dept Chem, Allentown, PA 18104 USA
[4] Juniata Coll, Dept Chem, Huntingdon, PA 16652 USA
[5] Restek Corp, Bellefonte, PA 16823 USA
关键词
D O I
10.1021/ac030367x
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The design, fabrication, and performance of gas chromatography columns etched in silicon substrates are described. Deep reactive-ion etching formed the 3-m-long, 150-mum-wide, 240-mum-deep rectangular cross section channels. A glass cover plate was anodically bonded to the remaining surface of the substrate forming the gastight channel. For some of the columns, the silicon channels were oxidized before the channels were sealed with the glass plates. Fused-silica capillary connecting tubes were sealed into ports on the edge of the 3.2-cm x 3.2-cm substrate chips. Dynamic coating was used to deposit a film of nonpolar dimethyl polysiloxane or moderately polar trifluoropropyhnethyl polysiloxane stationary phase. The columns were evaluated in a conventional benchtop GC instrument with split injection and flame ionization detection. Column efficiency was evaluated by the use of plots of height equivalent to a theoretical plate versus average carrier gas velocity using both hydrogen and air as carrier gases. The number of theoretical plates measured at the average carrier gas velocity giving the minimum plate height ranged from 4600 to 8200 plates for the dimethyl polysiloxane columns and from 3500 to 5500 plates for the trifluoropropyhnethyl polysiloxane columns. Minimum plate height was significantly smaller with air as carrier gas. For the nonpolar phase, the nonoxidized surface gave similar to1500 plates more than the oxidized surface for both carrier gases. For the polar phase, the oxidized surface gave similar to200 plates more than the nonoxidized surface. Isothermal chromatograms of a 20-component multifunctional mixture and temperature-programmed chromatograms of a normal alkane mixture are presented.
引用
收藏
页码:2629 / 2637
页数:9
相关论文
共 54 条
[1]  
Agah M, 2003, 12 INT C SOL STAT SE, P1339
[2]  
ASTLE A, 2002, P IMECE 02 2002 ASME
[3]  
BARTH PW, 1985, SCI AM, V248, P44
[4]   Sample filtration, concentration, and separation integrated on microfluidic devices [J].
Broyles, BS ;
Jacobson, SC ;
Ramsey, JM .
ANALYTICAL CHEMISTRY, 2003, 75 (11) :2761-2767
[5]   Dual-chemiresistor GC detector employing layer-protected metal nanocluster interfaces [J].
Cai, QY ;
Zellers, ET .
ANALYTICAL CHEMISTRY, 2002, 74 (14) :3533-3539
[6]   Gas chromatography [J].
Eiceman, GA ;
Gardea-Torresdey, J ;
Overton, E ;
Carney, K ;
Dorman, F .
ANALYTICAL CHEMISTRY, 2002, 74 (12) :2771-2780
[7]   SEPARATION VALUES AND THEIR UTILIZATION IN COLUMN CHARACTERIZATION .1. MEANING OF SEPARATION VALUES AND THEIR RELATIONSHIP TO OTHER CHROMATOGRAPHIC PARAMETERS [J].
ETTRE, LS .
CHROMATOGRAPHIA, 1975, 8 (06) :291-299
[8]   Lab-on-a-chip: A revolution in biological and medical sciences. [J].
Figeys, D ;
Pinto, D .
ANALYTICAL CHEMISTRY, 2000, 72 (09) :330A-335A
[9]  
Frye-Mason G, 2000, MESA MG, P229
[10]   LIQUID DISTRIBUTION ON GAS CHROMATOGRAPHIC SUPPORT - RELATIONSHIP TO PLATE HEIGHT [J].
GIDDINGS, JC .
ANALYTICAL CHEMISTRY, 1962, 34 (04) :458-&