Effect of rotational speed on the breakage of nickel-titanium rotary files

被引:32
作者
Dietz, DB [1 ]
Di Fiore, PM [1 ]
Bahcall, JK [1 ]
Lautenschlager, EP [1 ]
机构
[1] Northwestern Univ, Sch Dent, Chicago, IL 60611 USA
关键词
D O I
10.1097/00004770-200002000-00002
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Nickel-titanium 0.04-tapered rotary flies were evaluated for breakage at different rotational speeds in semicircular bovine bone simulated root canals of identical size and radius for each file size group tested. The bovine bone canals had a radius of curvature of 5 mm and a canal width equivalent to the Dl diameter of the file plus 0.04 mm, Profile instrument #3, #4, and #5 were tested at 150, 250, and 350 rpm. A contra-angle electric handpiece was mounted on an Instron machine that was set to deliver a constant downward speed of 5 mml min. The electric handpiece and Instron machine were activated until the files broke, The amount of file tip penetration into the semicircular bovine bone canal was measured in degrees with a protractor from a radiographic image taken of the file inside the bone model. Greater degrees of tip penetration indicated greater resistance to breakage. Statistical analysis was done and the results indicated that there was a significant difference for all file sizes in the extent of file tip penetration before breakage. In the rotation range between 150 and 350 rpm the greatest extent of penetration occurred at 150 rpm. This study concluded that 0.04 taper nickel-titanium rotary file breakage is less likely to occur if the files are rotated at lower speeds.
引用
收藏
页码:68 / 71
页数:4
相关论文
共 15 条
[1]  
COHEN S, 1998, PATHWAYS PULP, P255
[2]  
Council on Dental Materials and Devices, 1976, J AM DENT ASSOC, V93, P813, DOI 10.14219/jada.archive.1976.0087
[3]   A COMPARISON OF CANAL PREPARATION WITH NICKEL-TITANIUM AND STAINLESS-STEEL INSTRUMENTS [J].
ESPOSITO, PT ;
CUNNINGHAM, CJ .
JOURNAL OF ENDODONTICS, 1995, 21 (04) :173-176
[4]   A COMPARISON OF ROOT-CANAL PREPARATIONS USING NI-TI HAND, NI-TI ENGINE-DRIVEN, AND K-FLEX ENDODONTIC INSTRUMENTS [J].
GLOSSON, CR ;
HALLER, RH ;
DOVE, SB ;
DELRIO, CE .
JOURNAL OF ENDODONTICS, 1995, 21 (03) :146-151
[5]   AN EVALUATION OF NITINOL AND STAINLESS-STEEL FILES USED BY DENTAL STUDENTS DURING A LABORATORY PROFICIENCY EXAM [J].
HIMEL, VT ;
AHMED, KM ;
WOOD, DM ;
ALHADAINY, HA .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 1995, 79 (02) :232-237
[6]  
LUGASSY AA, 1968, THESIS U PENNSYLVANI
[7]   THE COMPARATIVE EFFICIENCY OF 4 TYPES OF ENDODONTIC INSTRUMENTS [J].
MACHIAN, GR ;
PETERS, DD ;
LORTON, L .
JOURNAL OF ENDODONTICS, 1982, 8 (09) :398-402
[8]   Torsional testing of the lightspeed nickel-titanium instrument system [J].
Marsicovetere, ES ;
Burgess, JO ;
Clement, DJ ;
delRio, CE .
JOURNAL OF ENDODONTICS, 1996, 22 (12) :681-684
[9]   CUTTING EFFICIENCY OF ENDODONTIC INSTRUMENTS .1. A QUANTITATIVE COMPARISON OF THE TIP AND FLUTED REGIONS [J].
MISERENDINO, LJ ;
MOSER, JB ;
HEUER, MA ;
OSETEK, EM .
JOURNAL OF ENDODONTICS, 1985, 11 (10) :435-441
[10]   CUTTING EFFICIENCY OF ENDODONTIC REAMERS [J].
OLIET, S ;
SORIN, SM .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1973, 36 (02) :243-252