Aerosol delivery from a metered-dose inhaler during mechanical ventilation - An in vitro model

被引:97
作者
Fink, JB [1 ]
Dhand, R [1 ]
Duarte, AG [1 ]
Jenne, JW [1 ]
Tobin, MJ [1 ]
机构
[1] VENTILATOR SUPPORT CTR,HINSDALE,IL
关键词
D O I
10.1164/ajrccm.154.2.8756810
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Successful bronchodilator therapy with a metered-dose inhaler (MDI) in intubated, mechanically ventilated patients requires adequate delivery of aerosol to the lower respiratory tract. We determined the effect of ventilator mode, inspiratory flow pattern, humidity, and spontaneous respiratory effort on albuterol delivery in a model of the trachea and bronchi. The model was ventilated through an endotracheal tube during controlled mechanical ventilation (CMV), assist control (At), pressure support (PS), and continuous positive airway pressure (CPAP), separately with a dry and humidified ventilator circuit. Delivery of albuterol administered by a MDI and spacer on filters placed at the ends of the bronchi was measured by spectrophotometry (246 nm). Under dry conditions and with a frequency of 10 breaths/min, albuterol delivery with CMV (VT, 800 mi; 30.3 +/- 3.4%), At (VT, 800 mi; 31.9 +/- 1.3%), PS 10 cm H2O (VT, 700 mi; 28.8 +/- 4.5%), or PS 20 cm H2O (VT, 800 ml; 30.9 +/- 1.8%) was lower than that observed with simulated spontaneous breaths with CPAP (VT, 800 mi; 39.2 +/- 1.4%) (p < 0.01 for all modes). Delivery was greater under dry (28.8 to 39%) than under humidified conditions (15.9 to 20.2%) (p < 0.005 in all modes). Albuterol delivery showed a linear correlation with both inspiratory time and duty cycle (r > 0.91). Lower respiratory tract delivery of aerosol from a MDI varied from 4.9 to 39.2%. We conclude that in addition to other known factors such as dose, type of spacer, and its position the technique of administering MDIs in mechanically ventilated patients markedly influences lower respiratory tract aerosol delivery.
引用
收藏
页码:382 / 387
页数:6
相关论文
共 32 条
[1]   THE DELIVERY OF THERAPEUTIC AEROSOLS THROUGH ENDOTRACHEAL-TUBES [J].
AHRENS, RC ;
RIES, RA ;
POPENDORF, W ;
WIESE, JA .
PEDIATRIC PULMONOLOGY, 1986, 2 (01) :19-26
[2]   METERED DOSE INHALER AEROSOL CHARACTERISTICS ARE AFFECTED BY THE ENDOTRACHEAL-TUBE ACTUATOR ADAPTER USED [J].
BISHOP, MJ ;
LARSON, RP ;
BUSCHMAN, DL .
ANESTHESIOLOGY, 1990, 73 (06) :1263-1265
[3]   DELIVERY EFFICIENCY OF METERED DOSE AEROSOLS GIVEN VIA ENDOTRACHEAL-TUBES [J].
CROGAN, SJ ;
BISHOP, MJ .
ANESTHESIOLOGY, 1989, 70 (06) :1008-1010
[4]   BRONCHODILATOR DELIVERY BY METERED-DOSE INHALER IN VENTILATOR-SUPPORTED PATIENTS [J].
DHAND, R ;
JUBRAN, A ;
TOBIN, MJ .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1995, 151 (06) :1827-1833
[5]  
DOLOVICH MB, 1977, AM REV RESPIR DIS, V115, P397
[6]  
Ebert J, 1992, Respir Care, V37, P862
[7]   EFFICIENCY OF BRONCHODILATOR AEROSOL DELIVERY TO THE LUNGS FROM THE METERED-DOSE INHALER IN MECHANICALLY VENTILATED PATIENTS - A STUDY COMPARING 4 DIFFERENT ACTUATOR DEVICES [J].
FULLER, HD ;
DOLOVICH, MB ;
TURPIE, FH ;
NEWHOUSE, MT .
CHEST, 1994, 105 (01) :214-218
[8]   PRESSURIZED AEROSOL VERSUS JET AEROSOL DELIVERY TO MECHANICALLY VENTILATED PATIENTS - COMPARISON OF DOSE TO THE LUNGS [J].
FULLER, HD ;
DOLOVICH, MB ;
POSMITUCK, G ;
PACK, WW ;
NEWHOUSE, MT .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1990, 141 (02) :440-444
[9]   AEROSOL DELIVERY DURING MECHANICAL VENTILATION - A PREDICTIVE INVITRO LUNG MODEL [J].
FULLER, HD ;
DOLOVICH, MB ;
CHAMBERS, C ;
NEWHOUSE, MT .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1992, 5 (04) :251-259
[10]  
GARNER SS, 1994, PHARMACOTHERAPY, V14, P210