Performance comparison of full-field digital mammography to screen-film mammography in clinical practice

被引:60
作者
Berns, EA
Hendrick, RE [1 ]
Cutter, GR
机构
[1] Northwestern Univ, Sch Med, Lynn Sage Comprehens Breast Ctr, Chicago, IL 60611 USA
[2] AMC Canc Res Ctr, Denver, CO USA
关键词
Clinical techniques - Control performance - Digital mammography - Digital mammography system - Full field digital mammography - Mean glandular dose - Performance comparison - Screen-film mammographies;
D O I
10.1118/1.1472497
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Results of acceptance testing 18 full-field digital mammography systems for clinical use and of conducting annual physics surveys of 38 screen-film mammography systems were compared in terms of exposure times, mean glandular breast doses, and image quality. These evaluations were made using the same test tools on all systems, with emphasis on assessing automatic exposure control performance and image quality on both digital and screen-film systems using clinical techniques. Survey results indicated that digital mammography systems performed similarly to screen-film systems in terms of exposure times and mean glandular doses for thin to intermediate breasts, but that digital mammography systems selected shorter exposure times and lower mean glandular doses for thicker breasts. For all breast thicknesses, digital mammography systems yielded mean contrast-detail scores higher than those for screen-film systems. For all breast thicknesses, the 18 digital mammography systems demonstrated less variance in terms of exposure times, mean glandular doses, and contrast-detail scores than did the 38 screen-film systems tested. These results indicate that the clinical use of digital mammography may generally improve image quality for equal or lower breast doses, while providing tighter control on exposures and image quality than screen-film mammography. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:830 / 834
页数:5
相关论文
共 9 条
  • [1] Botsco M, 1999, MAMMOGRAPHY QUALITY
  • [2] The effect of the antiscatter grid on full-field digital mammography phantom images
    Chakraborty, DP
    [J]. JOURNAL OF DIGITAL IMAGING, 1999, 12 (01) : 12 - 22
  • [3] Haus AG, 1997, BASICS FILM PROCESSI
  • [4] Optimizing techniques in screen-film mammography
    Hendrick, RE
    Berns, EA
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2000, 38 (04) : 701 - +
  • [5] Improvement in mammography quality control: 1987-1995
    Hendrick, RE
    Chrvala, CA
    Plott, CM
    Cutter, GR
    Jessop, NW
    Wilcox-Buchalla, P
    [J]. RADIOLOGY, 1998, 207 (03) : 663 - 668
  • [6] HENDRICK RE, 2001, ADV FILM PROCESSING, P115
  • [7] Comparison of full-field digital mammography with screen-film mammography for cancer detection: Results of 4,945 paired examinations
    Lewin, JM
    Hendrick, RE
    D'Orsi, CJ
    Isaacs, PK
    Moss, LJ
    Karellas, A
    Sisney, GA
    Kuni, CC
    Cutter, GR
    [J]. RADIOLOGY, 2001, 218 (03) : 873 - 880
  • [8] Miller RG, 1986, ANOVA BASICS APPL ST
  • [9] Full breast digital mammography with an amorphous silicon-based flat panel detector: Physical characteristics of a clinical prototype
    Vedantham, S
    Karellas, A
    Suryanarayanan, S
    Albagli, D
    Han, S
    Tkaczyk, EJ
    Landberg, CE
    Opsahl-Ong, B
    Granfors, PR
    Levis, I
    D'Orsi, CJ
    Hendrick, RE
    [J]. MEDICAL PHYSICS, 2000, 27 (03) : 558 - 567