Clinical evaluation of a computer-controlled pressure support mode

被引:93
作者
Dojat, M
Harf, A
Touchard, D
Lemaire, F
Brochard, L
机构
[1] Univ Paris 12, Hop Henri Mondor, Med Intens Care Unit, Dept Physiol,INSERM,U492,APHP, F-94010 Creteil, France
[2] Univ Paris 12, Hop Henri Mondor, Serv Reanimat Med, APHP, F-94010 Creteil, France
关键词
D O I
10.1164/ajrccm.161.4.9904064
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We have designed a computerized system providing closed-loop control of the level of pressure support ventilation (PSV). The system sets itself at the lowest level of PSV that maintains respiratory rate (RR), tidal volume (VT), and end-tidal CO2 pressure (PETCO2) within predetermined ranges defining acceptable ventilation (i.e., 12 < RR < 28 cycles/min, VT) 300 ml [> 250 if weight < 55 kg], and PETCO2 < 55 mm Hg [< 65 mm Hg if chronic CO2 retention]). Ten patients received computer-controlled (automatic) PSV and physician-controlled (standard) PSV, in random order, during 24 h for each mode. An estimation of occlusion pressure (P-0.1) was recorded continuously. The average time spent with acceptable ventilation as previously defined was 66 +/- 24% of the total ventilation time with standard PSV versus 93 +/- 8% with automatic PSV (p < 0.05), whereas the level of PSV was similar during the two periods (17 +/- 4 cm H2O versus 19 +/- 6 cm H2O). The time spent with an estimated Po., above 4 cm H2O was 34 +/- 35% of the standard PSV time versus only 11 +/- 17% of the automatic PSV time (p < 0.01). Automatic PSV increased the time spent within desired ventilation parameter ranges and apparently reduced periods of excessive workload.
引用
收藏
页码:1161 / 1166
页数:6
相关论文
共 28 条
[1]   P0.1 IS A USEFUL PARAMETER IN SETTING THE LEVEL OF PRESSURE SUPPORT VENTILATION [J].
ALBERTI, A ;
GALLO, F ;
FONGARO, A ;
VALENTI, S ;
ROSSI, A .
INTENSIVE CARE MEDICINE, 1995, 21 (07) :547-553
[2]   INSPIRATORY PRESSURE SUPPORT PREVENTS DIAPHRAGMATIC FATIGUE DURING WEANING FROM MECHANICAL VENTILATION [J].
BROCHARD, L ;
HARF, A ;
LORINO, H ;
LEMAIRE, F .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1989, 139 (02) :513-521
[3]  
BROCHARD L, 1994, PRINCIPLES PRACTICE, P239
[4]  
Brochard L, 1987, AM REV RESPIR DIS, V32, P1110
[5]  
CHAMBRIN MC, 1992, 14 IEEE ENG MED BIOL, P2419
[6]   Estimation of occlusion pressure during assisted ventilation in patients with intrinsic PEEP [J].
Conti, G ;
Cinnella, G ;
Barboni, E ;
Lemaire, F ;
Harf, A ;
Brochard, L .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1996, 154 (04) :907-912
[7]   EARLY PREDICTION OF SUCCESSFUL WEANING DURING PRESSURE SUPPORT VENTILATION IN CHRONIC OBSTRUCTIVE PULMONARY-DISEASE PATIENTS [J].
CONTI, G ;
DEBLASI, R ;
PELAIA, P ;
BENITO, S ;
ROCCO, M ;
ANTONELLI, M ;
BUFI, M ;
MATTIA, C ;
GASPARETTO, A .
CRITICAL CARE MEDICINE, 1992, 20 (03) :366-371
[8]   Evaluation of a knowledge-based system providing ventilatory management and decision for extubation [J].
Dojat, M ;
Harf, A ;
Touchard, D ;
Laforest, M ;
Lemaire, F ;
Brochard, L .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1996, 153 (03) :997-1004
[9]   A KNOWLEDGE-BASED SYSTEM FOR ASSISTED VENTILATION OF PATIENTS IN INTENSIVE-CARE UNITS [J].
DOJAT, M ;
BROCHARD, L ;
LEMAIRE, F ;
HARF, A .
INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1992, 9 (04) :239-250
[10]   NeoGanesh: a working system for the automated control of assisted ventilation in ICUs [J].
Dojat, M ;
Pachet, F ;
Guessoum, Z ;
Touchard, D ;
Harf, A ;
Brochard, L .
ARTIFICIAL INTELLIGENCE IN MEDICINE, 1997, 11 (02) :97-117