Improving Alarm Performance in the Medical Intensive Care Unit Using Delays and Clinical Context

被引:154
作者
Gorges, Matthias [1 ]
Markewitz, Boaz A. [2 ]
Westenskow, Dwayne R. [1 ]
机构
[1] Univ Utah, Dept Anesthesiol, Salt Lake City, UT 84132 USA
[2] Univ Utah, Dept Internal Med, Div Resp Crit Care & Occupat Pulm Med, Salt Lake City, UT 84132 USA
关键词
ANESTHESIA ALARMS; PULSE; RESPIRATION; FREQUENCY; BED;
D O I
10.1213/ane.0b013e31819bdfbb
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
INTRODUCTION: In an intensive care unit, alarms are used to call attention to a patient, to alert a change in the patient's physiology, or to warn of a failure in a medical device; however, up to 94% of the alarms are false. Our purpose in this study was to identify a means of reducing the number of false alarms. METHODS: An observer recorded time-stamped information of alarms and the presence of health care team members in the patient room; each alarm response was classified as effective (action taken within 5 min), ineffective (no response to the alarm), and ignored (alarm consciously ignored or actively silenced). RESULTS: During the 200-h study period, 1271 separate entries by an individual to the room being observed were recorded, 1214 alarms occurred and 2344 tasks were performed. On average, alarms occurred 6.07 times per hour and were active for 3.28 min per hour; 23% were effective, 36% were ineffective, and 41% were ignored. The median alarm duration was 1.7 s. A 14-s delay before alarm presentation would remove 50% of the ignored and ineffective alarms, and a 19-s delay would remove 67%. Suctioning, washing, repositioning, and oral care caused 152 ignored or ineffective ventilator alarms. DISCUSSION: Introducing a 19-s alarm delay and automatically detecting suctioning, repositioning, oral care, and washing could reduce the number of ineffective and ignored alarms from 934 to 274. More reliable alarms could elicit more timely response, reduce workload, reduce noise pollution, and potentially improve patient safety.
引用
收藏
页码:1546 / 1552
页数:7
相关论文
共 20 条
[1]   Context awareness in health care: A review [J].
Bricon-Souf, Nathalie ;
Newman, Conrad R. .
INTERNATIONAL JOURNAL OF MEDICAL INFORMATICS, 2007, 76 (01) :2-12
[2]   Multicentric study of monitoring alarms in the adult intensive care unit (ICU): a descriptive analysis [J].
Chambrin, MC ;
Ravaux, P ;
Calvelo-Aros, D ;
Jaborska, A ;
Chopin, C ;
Boniface, B .
INTENSIVE CARE MEDICINE, 1999, 25 (12) :1360-1366
[3]  
Christensen Martin, 2007, Nurs Crit Care, V12, P188, DOI 10.1111/j.1478-5153.2007.00229.x
[4]   Adapting to new technology in the operating room [J].
Cook, RI .
HUMAN FACTORS, 1996, 38 (04) :593-613
[5]  
DEY AK, 2000, THESIS GEORGIA I TEC, P188
[6]  
GOLDMAN JM, 2004, AUTOMATIC IDENTIFICA
[7]   Alarm algorithms in critical care monitoring [J].
Imhoff, M ;
Kuhls, S .
ANESTHESIA AND ANALGESIA, 2006, 102 (05) :1525-1537
[8]   A comparison of the failure times of pulse oximeters during blood pressure cuff-induced hypoperfusion in volunteers [J].
Kawagishi, T ;
Kanaya, N ;
Nakayama, M ;
Kurosawa, S ;
Namiki, A .
ANESTHESIA AND ANALGESIA, 2004, 99 (03) :793-796
[9]   FREQUENCY AND RELIABILITY OF ALARMS IN THE MONITORING OF CARDIAC POSTOPERATIVE-PATIENTS [J].
KOSKI, EMJ ;
MAKIVIRTA, A ;
SUKUVAARA, T ;
KARI, A .
INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1990, 7 (02) :129-133
[10]   CRYING WOLF - FALSE ALARMS IN A PEDIATRIC INTENSIVE-CARE UNIT [J].
LAWLESS, ST .