An evaluation of three glucose meter systems and their performance in relation to criteria of acceptability for neonatal specimens

被引:11
作者
St-Louis, P [1 ]
Ethier, J [1 ]
机构
[1] Hop St Justine, Dept Clin Biochem, Montreal, PQ H3T 1C5, Canada
关键词
glucose meter; evaluation; neonatal samples; precision; accuracy;
D O I
10.1016/S0009-8981(02)00160-2
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Prior to making a selection for our hospital, a pediatric institution, we deemed it necessary to evaluate, concurrently, three recently available glucose meter systems, claimed to be suitable for use with neonatal samples. Methods: Comparisons were with laboratory plasma analyses. Linearity and precision were also determined. Results: All meters gave linear responses. Precision determined using quality control material was acceptable. For an in-laboratory side-by-side evaluation, meter 1 showed a small bias but significant result scatter while meter 3 showed a negative bias; mean differences from reference (S.D.) were: -0.30 mmol/l (0.56), 0.06 mmol/l (0.39) and -0.49 mmol/l (0.35) for meters 1, 2 and 3, respectively. Clinical unit testing results gave mean differences from reference (S.D.) of: -0.19 mmol/l (0.56), 0.06 mmol/l (0.48) and -0.12 mmol/l (0.48) for meters 1, 2 and 3, respectively. Using 15% of reference as acceptability thresholds, 61%, 79% and 72% of results for meters 1, 2 and 3 respectively, were within limits. At 20%, the corresponding figures were 81%, 90% and 91%, respectively. All meters showed a sample-hematocrit effect with either negative (meters 1 and 3) or positive (meter 2) bias. Conclusions: Regardless of the performance criteria chosen, meter 1 had the worst performance while meter 2 was slightly better in overall than meter 3. Based on performance, general characteristics and user feedback, meter 2 was selected by us. In light of our results, we nonetheless suggest that performance of the meters tested is less than ideal, especially in the context of clinical utility in neonates. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:139 / 148
页数:10
相关论文
共 30 条
[1]  
*AM DIAB ASS, 1999, DIABETES CARE S1, V24, pS92
[2]  
[Anonymous], 1987, DIABETES CARE, V10, P95
[3]   THEORETICAL INVESTIGATION OF THE EFFECT OF FOULING ON THE PERFORMANCE OF A TUBE AND PLATE FIN HEAT-EXCHANGER [J].
BARROW, H ;
SHERWIN, K .
HEAT RECOVERY SYSTEMS & CHP, 1994, 14 (01) :1-5
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
Chance JJ, 1999, AM J CLIN PATHOL, V111, P547
[6]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[7]   Controversies regarding definition of neonatal hypoglycemia: Suggested operational thresholds [J].
Cornblath, M ;
Hawdon, JM ;
Williams, AF ;
Aynsley-Green, A ;
Ward-Platt, MP ;
Schwartz, R ;
Kalhan, SC .
PEDIATRICS, 2000, 105 (05) :1141-1145
[8]   Hypoglycemia in the neonate [J].
Cornblath, M ;
Ichord, R .
SEMINARS IN PERINATOLOGY, 2000, 24 (02) :136-149
[9]   Understanding error grid analysis [J].
Cox, DJ ;
GonderFrederick, LA ;
Kovatchev, BP ;
Julian, DM ;
Clarke, WL .
DIABETES CARE, 1997, 20 (06) :911-912
[10]  
Demers LM, 1999, CLIN CHEM, V45, P1578