Wavelength-tuning interferometry of intraocular distances

被引:147
作者
Lexer, F
Hitzenberger, CK
Fercher, AF
Kulhavy, M
机构
[1] Institute of Medical Physics, University of Vienna, Wien, A-1090
来源
APPLIED OPTICS | 1997年 / 36卷 / 25期
关键词
wavelength tuning; interferometry; wavelength-tuning interferometry; intraocular distances;
D O I
10.1364/AO.36.006548
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We describe basic principles of wavelength-tuning interferometry and demonstrate its application in ophthalmology. The advantage of this technique compared with conventional low-coherence interferometry ranging is the simultaneous measurement of the object structure without the need for a moving reference mirror. Shifting the wavelength of an external-cavity tunable laser diode causes intensity oscillations in the interference pattern of light beams remitted from the intraocular structure. A Fourier transform of the corresponding wave-number-dependent photodetector signal yields the distribution of the scattering potential along the light beam illuminating the eye. We use an external interferometer to linearize the wave-number axis. We obtain high resolution in a model eye by slow tuning over a wide wavelength range. With lower resolution we demonstrate the simultaneous measurement of anterior segment length, vitreous chamber depth, and axial eye length in human eyes in vivo with data-acquisition times in the millisecond range. (C) 1997 Optical Society of America.
引用
收藏
页码:6548 / 6553
页数:6
相关论文
共 38 条
  • [1] *AM NAT STAND I, 1986, 1361 ANSI Z
  • [2] BAUMGARTNER A, 1995, P SOC PHOTO-OPT INS, V2330, P146
  • [3] Bracewell R.N., 1965, The Fourier Transform and Its Applications
  • [4] Dandliker R., 1970, Optics Communications, V1, P323, DOI 10.1016/0030-4018(70)90032-5
  • [5] 3-DIMENSIONAL SENSING OF ROUGH SURFACES BY COHERENCE RADAR
    DRESEL, T
    HAUSLER, G
    VENZKE, H
    [J]. APPLIED OPTICS, 1992, 31 (07): : 919 - 925
  • [6] MEASUREMENT OF THE THICKNESS OF FUNDUS LAYERS BY PARTIAL COHERENCE TOMOGRAPHY
    DREXLER, W
    HITZENBERGER, CK
    SATTMANN, H
    FERCHER, AF
    [J]. OPTICAL ENGINEERING, 1995, 34 (03) : 701 - 710
  • [7] OPTICAL FREQUENCY-DOMAIN REFLECTOMETRY IN SINGLE-MODE FIBER
    EICKHOFF, W
    ULRICH, R
    [J]. APPLIED PHYSICS LETTERS, 1981, 39 (09) : 693 - 695
  • [8] Fercher A F, 1996, J Biomed Opt, V1, P157, DOI 10.1117/12.231361
  • [9] EYE-LENGTH MEASUREMENT BY INTERFEROMETRY WITH PARTIALLY COHERENT-LIGHT
    FERCHER, AF
    MENGEDOHT, K
    WERNER, W
    [J]. OPTICS LETTERS, 1988, 13 (03) : 186 - 188
  • [10] MEASUREMENT OF INTRAOCULAR DISTANCES BY BACKSCATTERING SPECTRAL INTERFEROMETRY
    FERCHER, AF
    HITZENBERGER, CK
    KAMP, G
    ELZAIAT, SY
    [J]. OPTICS COMMUNICATIONS, 1995, 117 (1-2) : 43 - 48