In vitro wear assessment of the Charite Artificial Disc according to ASTM recommendations

被引:29
作者
Serhan, Hassan A. [1 ]
Dooris, Andrew P. [1 ]
Parsons, Matthew L. [1 ]
Ares, Paul J. [1 ]
Gabriel, Stefan M. [1 ]
机构
[1] Johnson & Johnson, DePuy Spine, Raynham, MA 02767 USA
关键词
artificial disc; wear; biomechanics; lumbar;
D O I
10.1097/01.brs.0000228716.60863.ab
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Biomechanical laboratory research. Objective. To evaluate the potential for Ultra High Molecular Weight Polyethylene ( UHMWPE) wear debris from the Charite Artificial Disc. Summary of Background Data. Cases of osteolysis from artificial discs are extremely rare, but hip and knee studies demonstrate the osteolytic potential and clinical concern of UHMWPE wear debris. Standards for testing artificial discs continue to evolve, and there are few detailed reports of artificial disc wear characterizations. Methods. Implant assemblies were tested to 10 million cycles of +/- 7.5 degrees flexion-extension or +/- 7.5 degrees left/ right lateral bending, both with +/- 2 degrees axial rotation and 900 N to 1,850 N cyclic compression. Cores were weighed, measured, and photographed. Soak and loaded soak controls were used. Wear debris was analyzed via scanning electron microscopy and particle counters. Results. The average total wear of the implants was 0.11 and 0.13 mg per million cycles, before and after accounting for serum absorption, respectively. Total height loss was approximately 0.2 mm. Wear debris ranged from submicron to > 10 mu m in size. Conclusions. Under these test conditions, the Charite ' Artificial Disc produced minimal wear debris. Debris size and morphology tended to be similar to other CoCr-UHMWPE joints. More testing is necessary to evaluate the implants under a spectrum of loading conditions.
引用
收藏
页码:1900 / 1910
页数:11
相关论文
共 49 条
[1]  
*AM SOC TEST MAT, 2003, INT STAND TEST METH
[2]   The use of functional analysis in evaluating knee kinematics [J].
Andriacchi, TP ;
Dyrby, CO ;
Johnson, TS .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) :44-53
[3]  
Bargmann LS, 1999, CLIN ORTHOP RELAT R, P49
[4]   Effects of in vitro wear of machined and molded UHMWPE tibial inserts on TKR kinematics [J].
Benson, LC ;
DesJardins, JD ;
LaBerge, M .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2001, 58 (05) :496-504
[5]   A prospective, randomized, multicenter food and drug administration Investigational device exemptions study of lumbar total disc replacement with the CHARITE™ artificial disc versus lumbar fusion Part I:: Evaluation of clinical outcomes [J].
Blumenthal, S ;
McAfee, PC ;
Guyer, RD ;
Hochschuler, SH ;
Geisler, FH ;
Holt, RT ;
Garcia, R ;
Regan, JJ ;
Ohnmeiss, DD .
SPINE, 2005, 30 (14) :1565-1575
[6]  
Bragdon C R, 1996, Proc Inst Mech Eng H, V210, P157, DOI 10.1243/PIME_PROC_1996_210_408_02
[7]   Third-body wear of highly cross-linked polyethylene in a hip simulator [J].
Bragdon, CR ;
Jasty, M ;
Muratoglu, OK ;
O'Connor, DO ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2003, 18 (05) :553-561
[8]  
Bruce Martin R., 1998, Skeletal Tissue Mechanics
[9]   Influence of packaging conditions on the properties of gamma-irradiated UHMWPE following accelerated ageing and shelf ageing [J].
Buchanan, FJ ;
Sim, B ;
Downes, S .
BIOMATERIALS, 1999, 20 (09) :823-837
[10]  
BUSHELOW M, 2005, SPINE ARTHROPLASTY S