Multislice helical CT (MSCT) for mid-facial trauma: Optimization of parameters for scanning and reconstruction

被引:17
作者
Dammert, S
Funke, M
Merten, HA
Obernauer, S
Grabbe, E
机构
[1] Univ Gottingen, Zentrum Radiol, Rontgendiagnost Abt 1, D-37075 Gottingen, Germany
[2] Univ Gottingen, Abt Mund Kiefer & Gesichtschirurg, D-3400 Gottingen, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2002年 / 174卷 / 07期
关键词
multislice-CT; multiplanar reconstructions; facial trauma;
D O I
10.1055/s-2002-32682
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Purpose: To determine the optimal scan parameters in multislice helical CT (MSCT) of the facial bone complex for both axial scanning and multiplanar reconstructions. Material and Methods: An anthropomorphic skull phantom was examined with a MSCT. Axial scans were performed with continuously increasing collimations (4x1.25-4x2.5 mm), tube current (20200 mA) and table speeds (3.75 mm/rot. and 7.5 mm/rot.). Multiplanar reconstructions in coronal and parasagittal planes with different reconstruction increment and slice thickness were evaluated in terms of image noise, contour artifacts and visualisation of anatomical Structures. Results: The best image quality was obtained with a collimation of 4 x 1.25 mm and a table speed of 3.75 mm/rot. A reconstruction increment of 0.6 mm achieved the best time to image quality relation. With these parameters the bone structures were depicted in an optimal way without artifacts. The tube current could be reduced to 50 mA without significant loss of image quality. The optimized protocol was used for regular routine examinations in patients with facial trauma (n = 66). Conclusions: Low-dose MSCT using thin collimation, low table speed and small reconstruction increments provides excellent data for both axial images and multiplanar reconstructions in patients with facial trauma. An additional examination in coronal orientation is therefore no longer necessary.
引用
收藏
页码:874 / 879
页数:6
相关论文
共 14 条
[1]
Multislice spiral CT of the paranasal sinuses. [J].
Dammann, F ;
Bode, A ;
Heuschmid, M ;
Kopp, A ;
Georg, C ;
Pereira, PL ;
Claussen, CD .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (08) :701-706
[2]
GALANSKI M, 1998, GANZKORPERCOMPUTERTO, P12
[3]
Multi-slice helical CT: Scan and reconstruction [J].
Hu, H .
MEDICAL PHYSICS, 1999, 26 (01) :5-18
[4]
LOW-DOSE COMPUTED TOMOGRAPHIC IMAGING IN ORBITAL TRAUMA [J].
JACKSON, A ;
WHITEHOUSE, RW .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (788) :655-661
[5]
AN ANALYSIS OF FACIAL BONE-FRACTURES ON CT [J].
JEND, HH ;
JENDROSSMANN, JI ;
BORCHERS, D ;
HELLER, M .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NUKLEARMEDIZIN, 1982, 137 (04) :379-383
[6]
Treatment of frontobasal trauma and polytrauma [Behandlung von frontobasalen Traumen und Polytraumen.] [J].
Joos U. ;
Piffko J. ;
Meyer U. .
Mund-, Kiefer- und Gesichtschirurgie, 2001, 5 (2) :86-93
[7]
KALENDER W, 2000, COMPUTERTOMOGRAPHIE, P27
[8]
Multiplanar image reconstruction and 3D imaging using a musculoskeletal phantom; Conventional versus helical CT [J].
Kasales, CJ ;
Mauger, DT ;
Sefczek, RJ ;
Mahraj, R ;
Meilstrup, JW ;
VanHook, D ;
Patrone, SVM ;
Sefczek, DM ;
Eggli, KD ;
Wise, SW ;
Westacott, S ;
Hopper, KD .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1997, 21 (01) :162-169
[9]
Kopka L, 1999, RADIOLOGE, V39, P971, DOI 10.1007/s001170050590
[10]
COMPARISON OF HELICAL AND SERIAL CT WITH REGARD TO 3-DIMENSIONAL IMAGING OF MUSCULOSKELETAL ANATOMY [J].
NEY, DR ;
FISHMAN, EK ;
KAWASHIMA, A ;
ROBERTSON, DD ;
SCOTT, WW .
RADIOLOGY, 1992, 185 (03) :865-869