Elimination of the memory effects of gold, mercury and silver in inductively coupled plasma atomic emission spectroscopy

被引:37
作者
Chen, WC
Wee, P
Brindle, ID [1 ]
机构
[1] Brock Univ, Dept Chem, St Catharines, ON L2S 3A1, Canada
[2] Perkin Elmer Corp, Markham, ON L2R 8E5, Canada
关键词
D O I
10.1039/a908658g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The memory effects of mercury, gold and silver are eliminated by the use of sulfur containing complexing agents. In low concentrations (1 mu g ml(-1)) of mercury and gold, the addition of 2% L-cysteine gives wash-out times of less than 1 min for both elements. At higher concentrations of gold and mercury, and at low concentrations of silver, the reagent is less effective. Thiourea, at a concentration of 1%, eliminates the memory effects of silver, also giving wash-out times of approximately 1.5 min. The use of thiourea with higher concentrations (100 mu g ml(-1)) of mercury and gold delivers wash-out times of 1 min and 2 min, respectively. In a matrix similar to that found in many samples, the wash-out time for gold is less than 3.5 min and less than 2 min for mercury.
引用
收藏
页码:409 / 413
页数:5
相关论文
共 9 条
[1]  
BEARY E, COMMUNICATION
[2]   A COMPARISON OF CLOUD CHAMBERS FOR USE IN INDUCTIVELY COUPLED PLASMA NEBULIZATION SYSTEMS [J].
DALE, LS ;
BUCHANAN, SJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1986, 1 (01) :59-62
[3]   Use of gold as a routine and long term preservative for mercury in potable water, as determined by ICP-MS [J].
Fatemian, E ;
Allibone, J ;
Walker, PJ .
ANALYST, 1999, 124 (08) :1233-1236
[4]   Characterisation of memory effects and development of an effective wash protocol for the measurement of petrogenetically critical trace elements in geological samples by ICP-MS [J].
McGinnis, CE ;
Jain, JC ;
Neal, CR .
GEOSTANDARDS NEWSLETTER-THE JOURNAL OF GEOSTANDARDS AND GEOANALYSIS, 1997, 21 (02) :289-305
[5]   Simple direct method for the determination of total mercury levels in blood and urine and nitric acid digests of fish by inductively coupled plasma mass spectrometry [J].
Moreton, JA ;
Delves, HT .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (07) :659-665
[6]   Elements of ICPMS [J].
Newman, A .
ANALYTICAL CHEMISTRY, 1996, 68 (01) :A46-A51
[7]  
STROH A, 1993, OPTIMIZATION USE FLO
[8]  
TOTLAND M, COMMUNICATION