High-resolution mass spectrometry with a quadrupole operated in the fourth stability region

被引:17
作者
Chen, W [1 ]
Collings, BA [1 ]
Douglas, DJ [1 ]
机构
[1] Univ British Columbia, Dept Chem, Vancouver, BC V6T 1Z1, Canada
关键词
D O I
10.1021/ac990815u
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The properties of a quadrupole mass filter operated in the fourth stability region with Mathieu parameters (a,q) = (2 x 10(-3),21.3) have been investigated experimentally and theoretically. A resolution at half-height (R-1/2) of 13 900 has been obtained for 40 eV K-39(+) ions formed by thermal ionization, With ion energies of 750 eV or more, a resolution at half-height of up to 5000 can be obtained. Extrapolation of resolution data obtained with ions of 40 eV-4800 eV energy shows that unit resolution should be possible for ions with energies of similar to 10 keV, For low-energy ions (e.g., 75 eV), the transmission of the quadrupole operated in the fourth region is many orders of magnitude less than that of a quadrupole operated in the first or second stability regions. Acceptance calculations show this is a result of the combination of a much lower acceptance and strongly defocusing hinging fields. However, for higher ion energies, the transmission with operation in the fourth region is found to be comparable to that of a quadrupole operated at the same resolution in the second region. The implications for inductively coupled plasma mass spectrometry are briefly discussed.
引用
收藏
页码:540 / 545
页数:6
相关论文
共 16 条
[1]   High-resolution mass spectrometry with a multiple pass quadrupole mass analyzer [J].
Amad, MH ;
Houk, RS .
ANALYTICAL CHEMISTRY, 1998, 70 (23) :4885-4889
[2]  
Dawson P., 1995, AVS CLASSICS VACUUM
[3]   Ion source emittance influence on the transmission of a quadrupole operated in the second stability region [J].
Douglas, DJ ;
Konenkov, NV .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1998, 9 (10) :1074-1080
[4]   Inductively coupled plasma mass spectrometry with a quadrupole mass filter operated in the third stability region [J].
Du, ZH ;
Olney, TN ;
Douglas, DJ .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1997, 8 (12) :1230-1236
[5]   Elemental analysis with quadrupole mass filters operated in higher stability regions [J].
Du, ZH ;
Douglas, DJ ;
Konenkov, N .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (08) :1111-1119
[6]   Peak splitting with a quadrupole mass filter operated in the second stability region [J].
Du, ZH ;
Douglas, DJ ;
Konenkov, N .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1999, 10 (12) :1263-1270
[7]   NONSPECTROSCOPIC INTERELEMENT INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
GILLSON, GR ;
DOUGLAS, DJ ;
FULFORD, JE ;
HALLIGAN, KW ;
TANNER, SD .
ANALYTICAL CHEMISTRY, 1988, 60 (14) :1472-1474
[8]   AN IMPROVED MODEL OF THE FRINGING FIELDS OF A QUADRUPOLE MASS FILTER [J].
HUNTER, KL ;
MCINTOSH, BJ .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1989, 87 (02) :157-164
[9]   CHARACTERISTICS OF A QUADRUPOLE MASS FILTER IN THE SEPARATION MODE OF A FEW STABILITY REGIONS [J].
KONENKOV, NV ;
KRATENKO, VI .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1991, 108 (2-3) :115-136
[10]   Double-focusing mass spectrometers in ICPMS [J].
Moens, L ;
Jakubowski, N .
ANALYTICAL CHEMISTRY, 1998, 70 (07) :251A-256A