Performance of the enhanced Abbott AxSYM Cardiac Troponin I reagent in patients with heterophilic antibodies

被引:27
作者
Yeo, KTJ
Storm, CA
Li, Y
Jayne, JE
Brough, T
Quinn-Hall, KS
Fitzmaurice, TF
机构
[1] Dartmouth Med Sch, Dept Pathol, Lebanon, NH 03756 USA
[2] Dartmouth Hitchcock Med Ctr, Lebanon, NH 03756 USA
[3] Dartmouth Med Sch, Dept Med, Lebanon, NH 03756 USA
关键词
cardiac troponin I; heterophilic antibody;
D O I
10.1016/S0009-8981(99)00260-0
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The presence of heterophilic antibodies in the serum of a small subpopulation of individuals continues to cause false results for modem-day immunoassays. In order to determine the frequency of heterophilic antibody (HA)-related false positives within our population of positive cardiac troponin I (cTnI) patients, we assayed 200 samples using the original in-house cTnI assay (Abbott AxSYM) and the Bayer ACS:180 cTnI, which we had previously observed to be more effective at blocking HA interference. Four samples were identified as false positives based on discordant results between the two assays, as well as the correction of the false positives by treatment of the samples with heterophilic antibody blocking reagent (HBR). An 'enhanced' version of the AxSYM cTnI reagent was designed to greatly reduce or eliminate RA interference, and has now replaced the original reagents. The present study shows that the enhanced reagent significantly reduced or eliminated much of the HA interference. Comparative studies between the enhanced cTnI reagent and the original Abbott AxSYM cTnI reagent showed excellent correlation and equivalent diagnostic concordance, when HA samples were excluded from the analysis. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:13 / 23
页数:11
相关论文
共 24 条
[1]  
Apple FS, 1997, CLIN CHEM, V43, P2047
[2]  
BOERMAN OC, 1990, CLIN CHEM, V36, P888
[3]  
Bohner J, 1996, CLIN CHEM, V42, P2046
[4]  
BOSCATO LM, 1986, CLIN CHEM, V32, P1491
[5]  
BOSCATO LM, 1988, CLIN CHEM, V34, P27
[6]  
Cole LA, 1999, CLIN CHEM, V45, P313
[7]  
CSAKO G, 1988, CLIN CHEM, V34, P1481
[8]  
Després N, 1998, CLIN CHEM, V44, P440
[9]   Circulating cardiac troponin I in trauma patients without cardiac contusion [J].
Edouard, AR ;
Benoist, JF ;
Cosson, C ;
Mimoz, O ;
Legrand, A ;
Samii, K .
INTENSIVE CARE MEDICINE, 1998, 24 (06) :569-573
[10]  
Fitzmaurice TF, 1998, CLIN CHEM, V44, P2212