Are adult transcranial Doppler systems suitable for application in neonates?

被引:5
作者
Ipsiroglu, OS
Steck, J
Michel, E
Pessenhofer, H
Jorch, G
Pollak, A
机构
[1] FRAUNHOFER INST BIOMED ENGN, DEPT ULTRASOUND, ST INGBERT, GERMANY
[2] DEPT PAEDIAT, NEUKOLLN, GERMANY
[3] GRAZ UNIV, DEPT PHYSIOL, GRAZ, AUSTRIA
[4] UNIV MUNSTER, DEPT PAEDIAT, D-4400 MUNSTER, GERMANY
关键词
transcranial Doppler sonography; adaptation for neonates; ultrasound energy output; long term measurements of cerebral blood flow velocities;
D O I
10.1007/BF02282884
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Transcranial Doppler systems have not been available for monitoring of cerebral blood flow velocities in neonates because of potential hazardous effects of energy output from standard instruments developed for adult application. Aim of the study was to test commercially available transcranial Doppler instruments for their applicabiltiy in neonates and to develop guidelines for adaptation for safe neonatal use. Energy output of five commercially available transcranial Doppler instruments was measured with a hydrophone system and a radiation force balance. At the highest setting and at the nominal 10% attenuation level, five out of five and two out Of five instruments, respectively, had an energy output above the recommended limits. Power reduction was not linear in one instrument. Evaluation of safety devices (alarm, freeze mode, energy reduction facilities. display of energy values) showed that none of the tested instruments had an optimal setting fur safe neonatal application. Conclusion Commercially avail able transcranial Doppler Instruments should be evaluated critically for their energy output prior to their application in neonates. Special software for neonatal application of transcranial Doppler systems should be developed in order to provide extremly low energy output levels and devices indication of duration of Doppler insonation and energy output.
引用
收藏
页码:942 / 947
页数:6
相关论文
共 22 条
[1]   NON-INVASIVE TRANSCRANIAL DOPPLER ULTRASOUND RECORDING OF FLOW VELOCITY IN BASAL CEREBRAL-ARTERIES [J].
AASLID, R ;
MARKWALDER, TM ;
NORNES, H .
JOURNAL OF NEUROSURGERY, 1982, 57 (06) :769-774
[2]  
*AIUM, 1990, SAF CONS DIAGN ULTR
[3]   CYCLICAL VARIATIONS IN CEREBRAL BLOOD-FLOW VELOCITY [J].
ANTHONY, MY ;
EVANS, DH ;
LEVENE, MI .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1991, 66 (01) :12-16
[4]  
ANTICH PP, 1993, CALCIF TISSUE INT S1, V53, P157
[5]   EXPERIMENTAL VALIDATION OF PREDICTED TEMPERATURE RISES IN TISSUE-MIMICKING MATERIALS [J].
BACON, DR ;
SHAW, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (11) :1647-1659
[6]   IS DIAGNOSTIC ULTRASOUND SAFE - CURRENT INTERNATIONAL CONSENSUS ON THE THERMAL MECHANISM [J].
BARNETT, SB ;
KOSSOFF, G ;
EDWARDS, MJ .
MEDICAL JOURNAL OF AUSTRALIA, 1994, 160 (01) :33-37
[7]   EFFECTS OF LARYNGOSCOPY AND TRACHEAL INTUBATION ON CEREBRAL AND SYSTEMIC HEMODYNAMICS IN CHILDREN UNDER DIFFERENT PROTOCOLS OF ANESTHESIA [J].
BODE, H ;
UMMENHOFER, W ;
FREI, F .
EUROPEAN JOURNAL OF PEDIATRICS, 1993, 152 (11) :905-908
[8]   CEREBRAL BLOOD-FLOW VELOCITY PATTERNS DURING CARDIAC-SURGERY UTILIZING PROFOUND HYPOTHERMIA WITH LOW-FLOW CARDIOPULMONARY BYPASS OR CIRCULATORY ARREST IN NEONATES AND INFANTS [J].
BURROWS, FA ;
BISSONNETTE, B .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1993, 40 (04) :298-307
[9]   SILENT CEREBRAL MICROEMBOLI OCCURRING DURING CAROTID ANGIOGRAPHY - FREQUENCY AS DETERMINED WITH DOPPLER SONOGRAPHY [J].
DAGIRMANJIAN, A ;
DAVIS, DA ;
ROTHFUS, WE ;
DEEB, ZL ;
GOLDBERG, AL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 161 (05) :1037-1040
[10]   ON LINE CEREBRAL BLOOD-FLOW VELOCITY AND BLOOD-PRESSURE MEASUREMENT IN NEONATES - A NEW METHOD [J].
FENTON, AC ;
EVANS, DH ;
LEVENE, MI .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1990, 65 (01) :11-14