Risks of loosening of a prosthetic glenoid implanted in retroversion

被引:239
作者
Farron, Alain [1 ]
Terrier, Alexandre [1 ]
Buechler, Philippe [1 ]
机构
[1] Univ Lausanne, Hop Orthoped, CH-1005 Lausanne, Switzerland
关键词
D O I
10.1016/j.jse.2005.10.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Osteoarthritis of the shoulder is frequently associated with posterior glenoid wear, which may be difficult to correct during shoulder arthroplasty. This study was designed to evaluate the risks that a prosthetic glenoid implanted in retroversion will loosen. The scapula, the humerus, the rotator cuff, and a total shoulder prosthesis were reconstructed with a 3-dimensional finite element model. The glenoid was placed in 5 different angles of retroversion (0 degrees, 5 degrees, 10 degrees, 15 degrees, and 20 degrees). Location of the glenohumeral contact point, articular pressure, bone and cement stress, and micromotion around the glenoid implant were calculated during internal and external rotation. Glenoid retroversion induced a posterior displacement of the glenohumeral contact point during internal and external rotation, inducing a significant increase of stress within the cement mantel (+ 326%) and within the glenoid bone (+ 162%). Furthermore, a major increase of micromotion was measured of the bone-cement interface (+706%). According to this study, glenoid retroversion exceeding 10 degrees should be corrected during total shoulder arthroplasty. If the correction is impossible, not replacing the glenoid should be considered.
引用
收藏
页码:521 / 526
页数:6
相关论文
共 26 条
[1]
Bigliani L U, 1995, J Shoulder Elbow Surg, V4, P87
[2]
Buchler P, 2002, Comput Methods Biomech Biomed Engin, V5, P387, DOI 10.1080/1025584021000016843
[3]
A finite element model of the shoulder:: application to the comparison of normal and osteoarthritic joints [J].
Büchler, P ;
Ramaniraka, NA ;
Rakotomanana, LR ;
Iannotti, JP ;
Farron, A .
CLINICAL BIOMECHANICS, 2002, 17 (9-10) :630-639
[4]
Biphasic constitutive laws for biological interface evolution [J].
Buechler, P. ;
Pioletti, D. P. ;
Rakotomanana, L. R. .
BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2003, 1 (04) :239-249
[5]
COMPARISON OF THE FATIGUE CHARACTERISTICS OF CENTRIFUGED AND UNCENTRIFUGED SIMPLEX P-BONE CEMENT [J].
DAVIES, JP ;
BURKE, DW ;
OCONNOR, DO ;
HARRIS, WH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1987, 5 (03) :366-371
[6]
THE USE OF COMPUTERIZED-TOMOGRAPHY IN THE MEASUREMENT OF GLENOID VERSION [J].
FRIEDMAN, RJ ;
HAWTHORNE, KB ;
GENEZ, BM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (07) :1032-1037
[7]
Shoulder arthroplasty for advanced glenohumeral arthritis after anterior instability repair [J].
Green, A ;
Norris, TR .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2001, 10 (06) :539-545
[8]
Characteristics of unsatisfactory shoulder arthroplasties [J].
Hasan, SS ;
Leith, JM ;
Campbell, B ;
Kapil, R ;
Smith, KL ;
Matsen, FA .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2002, 11 (05) :431-441
[9]
Long-term results of total shoulder arthroplasty following bone-grafting of the glenoid [J].
Hill, JM ;
Norris, TR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (06) :877-883
[10]
The effects of glenoid component alignment variations on cement mantle stresses in total shoulder arthroplasty [J].
Hopkins, AR ;
Hansen, UN ;
Amis, AA ;
Emery, R .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2004, 13 (06) :668-675