US with extended field of view: Phantom-tested accuracy of distance measurements

被引:43
作者
Fornage, BD
Atkinson, EN
Nock, LF
Jones, PH
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Diagnost Radiol, Houston, TX 77030 USA
[2] Univ Texas, MD Anderson Canc Ctr, Dept Biomath, Houston, TX 77030 USA
[3] Siemans Ultrasound Grp, Issaquah, WA USA
关键词
images; display; processing; ultrasound; (US); experimental; technology;
D O I
10.1148/radiology.214.2.r00fe20579
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To evaluate the accuracy of distance measurements obtained with the extended-field-of-view (FOV) software of a commercially available ultrasonographic scanner, two custom-designed phantoms that allowed scanning of flat and curved surfaces were used. Five hundred forty measurements of various known distances in the phantoms were made by three examiners using various transducers. Although minor differences were observed between operators and transducers, 99.4% (537 of 540) of the distance measurements were accurate within plus or minus 4%. This extended-FOV technology provides accurate measurements of large objects in vitro.
引用
收藏
页码:579 / 584
页数:6
相关论文
共 3 条
  • [1] Korger K, 1998, ULTRASOUND MED BIOL, V24, P1125, DOI 10.1016/S0301-5629(98)00066-0
  • [2] US extended-field-of-view imaging technology
    Weng, L
    Tirumalai, AP
    Lowery, CM
    Nock, LF
    Gustafson, DE
    VonBehren, PL
    Kim, JH
    [J]. RADIOLOGY, 1997, 203 (03) : 877 - 880
  • [3] WENG L, 1996, Patent No. 5575286