Refreshable mercury film silver based electrode for determination of chromium(VI) using catalytic adsorptive stripping voltammetry

被引:101
作者
Bas, B [1 ]
机构
[1] AGH Univ Sci & Technol, Fac Mat Sci & Ceram, PL-30059 Krakow, Poland
关键词
chromium speciation; water analysis; voltammetry; silver amalgams; mercury film electrodes; hanging mercury drop electrode;
D O I
10.1016/j.aca.2006.04.013
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The refreshable mercury film silver based electrode Hg(Ag)FE applied for determination of Cr(VI) traces using catalytic adsorptive striping voltammetry (CAdSV) will be presented. The film electrode is characterized by its very good surface reproducibility (not less than 2%) and long-term stability (1500-2000 measurement cycles). The mechanical refreshing of mercury film is realized in the specially constructed device, in a time shorter than 1-2 s. In the paper, it will be proved that a mechanically weak hanging mercury drop electrode (HMDE) may be substituted by mercury film Hg(Ag)FE electrode with a surface area adjustable from 1.5 to 12 mm(2). For the electrode surface 4 mm(2) the detection limit obtained for Cr(VI) was 0.19 nM, while the linearity range measured for a 20 s accumulation time was between 0.5 and 50 nM. The relative standard deviation (R.S.D.) in determination of Cr(VI) varied from 1 to 5%. The influence of the excess of Cr(HI) on determination of Cr(VI) was analyzed using samples from the Dobczyce reservoir spiked with known amounts of Cr(VI) and Cr(III). (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:195 / 201
页数:7
相关论文
共 23 条
[1]  
Bas B, 1997, Polish Patent, Patent No. [P-319984, 319984]
[2]  
BAS B, 2002, ELECTROANAL, V15, P14
[3]  
Bobrowski A, 2000, ELECTROANAL, V12, P1177, DOI 10.1002/1521-4109(200010)12:15<1177::AID-ELAN1177>3.0.CO
[4]  
2-U
[5]   Chromium speciation study in polluted waters using catalytic adsorptive stripping voltammetry and tangential flow filtration [J].
Bobrowski, A ;
Bas, B ;
Dominik, J ;
Niewiara, E ;
Szalinska, E ;
Vignati, D ;
Zarebski, J .
TALANTA, 2004, 63 (04) :1003-1012
[6]  
Bobrowski A, 2004, ACTA CHIM SLOV, V51, P77
[7]   THE DETERMINATION OF THE CHROMIUM SPECIATION IN SEA-WATER USING CATALYTIC CATHODIC STRIPPING VOLTAMMETRY [J].
BOUSSEMART, M ;
VAN DEN BERG, CMG ;
GHADDAF, M .
ANALYTICA CHIMICA ACTA, 1992, 262 (01) :103-115
[8]   STRIPPING VOLTAMMETRY OF CHROMIUM(VI) AT A POLY(4-VINYLPYRIDINE)-COATED PLATINUM-ELECTRODE [J].
COX, JA ;
KULESZA, PJ .
ANALYTICA CHIMICA ACTA, 1983, 154 (NOV) :71-78
[9]   Partitioning of chromium (VI) and chromium (III) between dissolved and colloidal forms in a stream and reservoir contaminated with tannery waste water [J].
Dominik, J ;
Bas, B ;
Bobrowski, A ;
Dworak, T ;
Koukal, B ;
Niewiara, E ;
de Abreu, MHP ;
Rossé, P ;
Szalinska, E ;
Vignati, D .
JOURNAL DE PHYSIQUE IV, 2003, 107 :385-388
[10]   DETERMINATION OF TRACE AMOUNTS OF CHROMIUM(VI) IN WATER BY ELECTROCHEMICAL METHODS [J].
ELLEOUET, C ;
QUENTEL, F ;
MADEC, C .
ANALYTICA CHIMICA ACTA, 1992, 257 (02) :301-308