Hot-wire amperometric monitoring of flowing streams

被引:23
作者
Wang, J [1 ]
Jasinski, M
Flechsig, GU
Grundler, P
Tian, BM
机构
[1] New Mexico State Univ, Dept Chem & Biochem, Las Cruces, NM 88003 USA
[2] Univ Rostock, Fachbereich Chem, D-18051 Rostock, Germany
关键词
electrochemical detection; flow analysis; heat pulse; hot wire electrochemistry; hydrodynamic modulation;
D O I
10.1016/S0039-9140(99)00224-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper describes the design of a hot-wire electrochemical flow detector, and the advantages accrued from the effects of locally increased temperature, mainly thermally induced convection, upon the amperometric monitoring of flowing streams. A new hydrodynamic modulation voltammetric approach is presented, in which the solution flow rate remains constant while the temperature of the working electrode is modulated. Factors influencing the response, including the flow rate, temperature pulse, or applied potential, have been investigated. The hot-wire operation results also in a significant enhancement of the flow injection amperometric response. The minimal flow rate dependence observed with the heated electrode should benefit the on-line monitoring of streams with fluctuated natural convection, as well as various in-situ remote sensing applications. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1205 / 1210
页数:6
相关论文
共 12 条