Internet-based home asthma telemonitoring - Can patients handle the technology?

被引:149
作者
Finkelstein, J
Cabrera, MR
Hripcsak, G
机构
[1] Columbia Univ, Dept Med Informat, New York, NY USA
[2] Columbia Presbyterian Med Ctr, Div Pulm Med, New York, NY 10032 USA
关键词
asthma; peak expiratory flow; self-management; spirometry; telemonitoring;
D O I
10.1378/chest.117.1.148
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Study objectice: To evaluate the validity of spirometry self-testing during home telemonitoring and to assess the acceptance of an Internet-based home asthma telemonitoring system by asthma patients. Design: We studied an Internet-based telemonitoring system that collected spirometry. data and symptom reports from asthma patients' homes for review by physicians in the medical center's clinical information system. After a 40-min training session, patients completed an electronic diary and performed spirometry testing twice daily on their own from their homes for 3 weeks. A medical professional visited each patient by the end of the third neck of monitoring, 10 to 40 min after the patient had performed self-testing, and asked the patient to perform the spirometry test again under his supervision, We evaluated the validity of self-testing and surveyed the patients attitude toward the technology using a standardized questionnaire, Setting: Telemonitoring was conducted in patients' homes in a low-income inner city area, Patients: Thirty-one consecutive asthma patients without regard to computer experience. Measurement and results: Thirty-one asthma patients completed 3 weeks of monitoring, A paired t test showed no difference between unsupervised and supervised home spirometry self-testing, The variability of FVC (4.1%), FEV1, (3.7%), peak expiratory flow (7.9%), and other spirometric indexes in our study was similar to the within-subject variability reported by other researchers. Despite the fact that the majority of the patients (71%) had no computer experience, they indicated that the self-testing was "not complicated at all" or only "slightly complicated." The majority of patients (87.1%) were strongly interested in using home asthma telemonitoring in the future. Conclusions: Spirometry self-testing by asthma patients during telemonitoring is valid and comparable to those tests collected under the supervision of a trained medical professional, Internet-based home asthma telemonitoring can be successfully implemented in a group of patients with no computer background.
引用
收藏
页码:148 / 155
页数:8
相关论文
共 33 条
[1]  
AFSCHRIFT M, 1969, AM REV RESPIR DIS, V100, P147
[3]  
[Anonymous], 1997, EVALUATION METHODS M
[4]   A SELF MANAGEMENT PLAN IN THE TREATMENT OF ADULT ASTHMA [J].
BEASLEY, R ;
CUSHLEY, M ;
HOLGATE, ST .
THORAX, 1989, 44 (03) :200-204
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   THE IMPORTANCE OF SPIROMETRY IN THE ASSESSMENT OF CHILDHOOD ASTHMA [J].
BYE, MR ;
KERSTEIN, D ;
BARSH, E .
AMERICAN JOURNAL OF DISEASES OF CHILDREN, 1992, 146 (08) :977-978
[7]   Residual abnormalities of pulmonary function in asymptomatic young adult asthmatics with childhood-onset asthma [J].
Chiang, CH ;
Hsu, K .
JOURNAL OF ASTHMA, 1997, 34 (01) :15-21
[8]   INTRASUBJECT VARIABILITY OF MAXIMAL EXPIRATORY FLOW VOLUME CURVE [J].
COCHRANE, GM ;
PRIETO, F ;
CLARK, TJH .
THORAX, 1977, 32 (02) :171-176
[9]   ESTIMATION OF THE COEFFICIENT OF VARIATION FROM LABORATORY ANALYSIS OF SPLIT SPECIMENS FOR QUALITY-CONTROL IN CLINICAL-TRIALS [J].
CONNETT, JE ;
LEE, WW .
CONTROLLED CLINICAL TRIALS, 1990, 11 (01) :24-36
[10]   VARIABILITY OF PULMONARY-FUNCTION TESTS IN CYSTIC-FIBROSIS [J].
COOPER, PJ ;
ROBERTSON, CF ;
HUDSON, IL ;
PHELAN, PD .
PEDIATRIC PULMONOLOGY, 1990, 8 (01) :16-22