Difference in blood pressure readings with mercury and automated devices: impact on hypertension prevalence estimates in Dar es Salaam, Tanzania

被引:9
作者
Chiolero, Arnaud
Gervasoni, Jean-Pierre
Rwebogora, Anne
Balampama, Marianna
Paccaud, Fred
Bovet, Pascal
机构
[1] Univ Lausanne, Inst Social & Prevent Med, CH-1005 Lausanne, Switzerland
[2] Minist Reg Adm & Local Govt, Dar Es Salaam, Tanzania
关键词
blood pressure measurement; hypertension; Tanzania; validation protocol;
D O I
10.1007/s10654-006-9015-z
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Objectives: ( 1) To compare blood pressure ( BP) readings with an automated arm cuff oscillometric device ( AutoBP) to readings with a mercury sphygmomanometer ( HgBP) and ( 2) to evaluate the impact on the prevalence of hypertension ( HBP) in a population-based survey. Methods: (1) In a convenience sample (''Comparison Study''), we measured BP with both AutoBP (Visomat (R) OZ2) and HgBP and we modeled BP difference ( Delta BP = HgBP) AutoBP) with multiple regression analysis. ( 2) Using DBP, we calculated HgBP in a survey previously conducted in Dar es Salaam (''Population Survey'') in which BP was measured with the automatic device Visomat((R)) OZ2 and we compared the prevalence of HBP (>= 140/90 mmHg or treatment). Results: In the Comparison Study ( 404 subjects aged 25 - 64), systolic/diastolic BP was higher by 4.4/4.7 mmHg ( SE: 0.4/0.3) with HgBP than AutoBP. The prevalence of HBP was 42% with HgBP and 36% with AutoBP ( relative difference of 14%). DBP was associated with age, BP and arm circumference. In the Population Survey ( 9.254 subjects aged 25 - 64), the prevalence of HBP was 17% with calculated HgBP and 14% with AutoBP ( relative difference of 20%). Conclusion: A small systematic bias in BP readings between two different devices had large impact on hypertension prevalence estimates. This suggests that automated devices used in epidemiological studies should be validated with particular care.
引用
收藏
页码:427 / 433
页数:7
相关论文
共 27 条
  • [1] Ahmad O, 2000, GPE DISCUSSION PAPER, V31
  • [2] [Anonymous], EL AUT SPHYGM
  • [3] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [4] Distribution of blood pressure, body mass index and smoking habits in the urban population of Dar es Salaam, Tanzania, and associations with socioeconomic status
    Bovet, P
    Ross, AG
    Gervasoni, JP
    Mkamba, M
    Mtasiwa, DM
    Lengeler, C
    Whiting, D
    Paccaud, F
    [J]. INTERNATIONAL JOURNAL OF EPIDEMIOLOGY, 2002, 31 (01) : 240 - 247
  • [5] Assessing the prevalence of hypertension in populations: are we doing it right?
    Bovet, P
    Gervasoni, JP
    Ross, AG
    Mkamba, M
    Mtasiwa, DM
    Lengeler, C
    Burnier, M
    Paccaud, F
    [J]. JOURNAL OF HYPERTENSION, 2003, 21 (03) : 509 - 517
  • [6] BRUCE NG, 1988, J HYPERTENS, V6, P375
  • [7] Discordant prevalence of hypertension using two different automated blood pressure measurement devices: a population-based study in Dar es Salaam (Tanzania)
    Chiolero, A
    Gervason, JP
    Rwebogora, A
    Mkamba, M
    Waeber, B
    Paccaud, F
    Burnier, M
    Bovet, P
    [J]. BLOOD PRESSURE MONITORING, 2004, 9 (02) : 59 - 64
  • [8] Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure
    Chobanian, AV
    Bakris, GL
    Black, HR
    Cushman, WC
    Green, LA
    Izzo, JL
    Jones, DW
    Materson, BJ
    Oparil, S
    Wright, JT
    Roccella, EJ
    [J]. HYPERTENSION, 2003, 42 (06) : 1206 - 1252
  • [9] Cooper, 1997, Blood Press Monit, V2, P35
  • [10] Dieterle T, 1998, Blood Press Monit, V3, P339