Sensitive determination of ethanol in low-alcohol samples by ion-exclusion chromatography with EC detection using a ruthenium-based inorganic film electrode

被引:5
作者
Cataldi, TRI
Centonze, D
Desimoni, E
机构
[1] Dipartimento di Chimica, Univ. degli Studi della Basilicata, 85-85100 Potenza, Via N. Sauro
关键词
D O I
10.1016/0308-8146(95)00060-7
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
Ethanol in real matrices was determined by ion-exclusion chromatography (IEC) with electrochemical detection at a glassy carbon electrode coated with conducting films of mixed-valent ruthenium oxide cross-linked with ruthenium cyanide (mvRuO-RuCN). These films exhibit very high and persistent elecrocatalytic behavior toward oxidation of hydroxyl-containing compounds, and offer attractive features for applications as electrocatalytic amperometric sensor in acidic media. Real samples such as vinegar and low-alcohol beers were simply diluted with mobile phase (1:10), filtered and injected without extensive sample treatment. A polymeric-based hydrogen-form ion-exclusion column was chosen for the chromatographic separations using 25-mM sulphuric acid as the mobile phase. The response was linear for sample solutions containing ethanol up to 0.1 M with a dynamic range of more than three orders of magnitude.
引用
收藏
页码:17 / 21
页数:5