MRI-induced heating of selected thin wire metallic implants - laboratory and computational studies - findings and new questions raised

被引:48
作者
Bassen, H [1 ]
Kainz, W [1 ]
Mendoza, G [1 ]
Kellom, T [1 ]
机构
[1] US FDA, Ctr Devices & Radiol Hlth, Rockville, MD 20852 USA
关键词
MRI; magnetic resonance imaging; medical implant; heating; safety; RF; radio frequency;
D O I
10.1080/13645700600640931
中图分类号
R61 [外科手术学];
学科分类号
摘要
We performed experiments and computer modeling of heating of a cardiovascular stent and a straight, thin wire by RF fields in a 1.5 T MRI birdcage coil at 64 MHz. We used ASTM F2182-02a standard and normalized results to 4 W/kg whole body average. We used a rectangular saline-gel filled phantom and a coiled, double stent ( Intracoil by ev3 Inc) 11 cm long. The stent had thin electrical insulation except for bare ends ( simulating drug eluting coating). The stent and phantom were placed close to the wall of the RF Coil and had approximately 0.5 degrees C initial temperature rise at the ends ( local SAR=320 W/kg). We exposed a wire ( 24.1 cm, 0.5 mm diameter) with 0.5 mm insulation and saw an 8.6 degrees C temperature rise ( local SAR=5680 W/kg) at the bare ends. All heating was within 1 mm(3) of the ends, so the position of our fiber optic temperature probe was critical for repeatability. Our computational study used finite difference time domain software with a thermodynamics solver. We modeled a coiled bare-wire stent as a spiral with a rectangular cross section and found a maximum increase of 0.05 degrees C induced at the tips for plane wave exposures. A maximum local SAR of up to 200 W/kg occurred in a volume of only 8 x 10(-3) mm. We developed improved computational exposure sources - optimized birdcage coils and quasi-MRI fields that may eliminate the need to model an RF coil. We learned that local ( point) SAR ( initial linear temperature rise) is the most reliable indicator of the maximum heating of an implant. Local SAR depends greatly on implant length, insulation and shape, and position in the MRI coil. Accurate heating must be measured with sensors or software having millimeter resolution. Many commercially available fiber optic temperature probes do meet this requirement.
引用
收藏
页码:76 / 84
页数:9
相关论文
共 13 条
[1]  
*AAMI, PC692000 ANSIAAMI
[2]  
[Anonymous], 60601233 IEC
[3]  
*ASTM, F218202A ASTM
[4]  
CALCAGNINI, COMMUNICATION
[5]   The dielectric properties of biological tissues .3. Parametric models for the dielectric spectrum of tissues [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2271-2293
[6]   Safety of metallic implants in magnetic resonance imaging [J].
Ho, HS .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (04) :472-477
[7]  
National Electrical Manufacturers Association (NEMA), 1993, 81993 NEMA MS
[8]  
Nitz WR, 2001, J MAGN RESON IMAGING, V13, P105, DOI 10.1002/1522-2586(200101)13:1<105::AID-JMRI1016>3.0.CO
[9]  
2-0
[10]   Neurostimulation systems for deep brain stimulation: In vitro evaluation of magnetic resonance imaging-related heating at 1.5 tesla [J].
Rezai, AR ;
Finelli, D ;
Nyenhuis, JA ;
Hrdlicka, G ;
Tkach, J ;
Sharan, A ;
Rugieri, P ;
Stypulkowski, PH ;
Shellock, FG .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 15 (03) :241-250