Experience of clinical use of the Odstock dropped foot stimulator

被引:71
作者
Burridge, J
Taylor, P
Hagan, S
Swain, I
机构
[1] Dept. of Med. Phys. and Biomed. Eng., Salisbury District Hospital, Salisbury, Wiltshire
[2] Dept. of Med. Phys. and Biomed. Eng., Salisbury District Hospital, Salisbury, Wiltshire
关键词
Odstock dropped foot stimulator; common peroneal nerve stimulation;
D O I
10.1111/j.1525-1594.1997.tb04662.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Odstock dropped foot stimulator (ODFS) is a simple functional electrical stimulation (FES) device for the correction of dropped foot. Improved reliability, fine control of stimulation parameters, and careful application and follow-up have led to 86% compliance. Data on 56 patients (50 patients with hemiplegia, 5 patients with multiple sclerosis, and 1 patient with spinal cord injury) who have used the system for between 6 and 18 months are presented and show a statistically significant increase in walking speed with the stimulator at 3 months of 14% (p < 0.001); decreased effort of walking, measured as physiological cost index (PCI), of 37% (p < 0.001); and statistically significant improvement in functional mobility tests and questionnaires. No statistically significant carryover was seen although 3 patients had sufficient improvement in active ankle control and gait parameters to no longer need the stimulator. Six patients who used the stimulator all day every day had a problem with skin irritation, which we have not yet been able to solve. Two patients discontinued use after experiencing increased spasticity in the calf.
引用
收藏
页码:254 / 260
页数:7
相关论文
共 4 条
[1]   Peroneal stimulator: Evaluation for the correction of spastic drop foot in hemiplegia [J].
Granat, MH ;
Maxwell, DJ ;
Ferguson, ACB ;
Lees, KR ;
Barbenel, JC .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1996, 77 (01) :19-24
[2]  
GRANCANIN F, 1984, INT REHABIL, V6, P25
[3]  
Liberson W., 1961, ARCH PHYS MED REHAB, V42, P202
[4]  
MERLETTI R, 1979, SCAND J REHABIL MED, V11, P111