Infrared spectroscopy in aqueous solution: Difficulties and accuracy of water subtraction

被引:87
作者
Rahmelow, K [1 ]
Hubner, W [1 ]
机构
[1] UNIV FREIBURG,INST PHYS CHEM,D-79104 FREIBURG,GERMANY
关键词
infrared spectroscopy; Fourier transform; water subtraction; aqueous solution; spectroscopic accuracy;
D O I
10.1366/0003702971940080
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The long-term reproducibility of a set of water spectra in the infrared region with cell thicknesses of less than 10 mu m was tested. With the application of a computer routine for subtraction in the range from 2600 to 1100 cm(-1) and the use of additional phase and nonlinearity corrections, an accuracy better than one milliabsorbance unit (mAU) can he achieved, if the temperature is kept constant within a tenth of a degree or if a temperature correction factor is implemented in the subtraction routine. The subtraction of water from an aqueous protein solution reduces the spectral range for a correction to 2300-1800 cm(-1). In this region the protein absorption, as derived from dry protein films, is constant, and because of the association band at 2125 cm(-1) water exhibits good subtraction properties. However, the reduction of the spectral range for the water subtraction leads to an increase in the residual error to 12 mAU at 1650 cm(-1). This standard deviation can be reduced to 6-7 mAU by applying a second subtraction range from 4000 to 3650 cm(-1). A further reduction of this error to 3-5 mAU is obtained by implementing a constant ratio of the integrated intensities of the amide I and amide II bands in the procedure of subtracting water from aqueous protein solutions. The remaining inaccuracy is mainly caused by uncompensated phase error.
引用
收藏
页码:160 / 170
页数:11
相关论文
共 12 条
[1]  
Born M., 1959, Principles of Optics
[2]   APPLICATION OF FOURIER-TRANSFORM INFRARED TRANSMISSION SPECTROSCOPY TO THE STUDY OF MODEL AND NATURAL MEMBRANES [J].
CAMERON, DG ;
CASAL, HL ;
MANTSCH, HH .
JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 1979, 1 (01) :21-36
[3]   PROTEIN SECONDARY STRUCTURES IN WATER FROM 2ND-DERIVATIVE AMIDE-I INFRARED-SPECTRA [J].
DONG, A ;
HUANG, P ;
CAUGHEY, WS .
BIOCHEMISTRY, 1990, 29 (13) :3303-3308
[4]   ON THE SPECTRAL SUBTRACTION OF WATER FROM THE FT-IR SPECTRA OF AQUEOUS-SOLUTIONS OF PROTEINS [J].
DOUSSEAU, F ;
THERRIEN, M ;
PEZOLET, M .
APPLIED SPECTROSCOPY, 1989, 43 (03) :538-542
[5]  
GRIFFITHS P, 1986, FOURIER TRANSFORM IN
[6]   STRUCTURE OF LIQUID WATER DETERMINED FROM INFRARED TEMPERATURE PROFILING AND EVOLUTIONARY CURVE RESOLUTION [J].
LIBNAU, FO ;
TOFT, J ;
CHRISTY, AA ;
KVALHEIM, OM .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1994, 116 (18) :8311-8316
[7]   SPECTRA OF WATER IN THE NEARINFRARED AND MIDINFRARED REGION [J].
LIBNAU, FO ;
KVALHEIM, OM ;
CHRISTY, AA ;
TOFT, J .
VIBRATIONAL SPECTROSCOPY, 1994, 7 (03) :243-254
[8]   A REDETERMINATION OF SOME OPTICAL PROPERTIES OF CALCIUM FLUORIDE [J].
MALITSON, IH .
APPLIED OPTICS, 1963, 2 (11) :1103-1107
[9]   AN ALGORITHM FOR THE REPRODUCIBLE SPECTRAL SUBTRACTION OF WATER FROM THE FT-IR SPECTRA OF PROTEINS IN DILUTE-SOLUTIONS AND ADSORBED MONOLAYERS [J].
POWELL, JR ;
WASACZ, FM ;
JAKOBSEN, RJ .
APPLIED SPECTROSCOPY, 1986, 40 (03) :339-344
[10]  
RAHMELOW K, UNPUB APPL OPT