On the statistical significance of electrophysiological steady-state responses

被引:143
作者
Meigen, T
Bach, M
机构
[1] Univ Wurzburg, Augenklin, Elektrophysiol Lab, D-97080 Wurzburg, Germany
[2] Univ Freiburg, Augenklin, D-79106 Freiburg, Germany
关键词
ERG; signal-to-noise ratio; statistical significance; steady-state; VEP;
D O I
10.1023/A:1002097208337
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Steady-state stimulation is a useful paradigm in many physiologic and clinical situations, for ERG, Pattern-ERG and VEP. One of the advantages is the easy evaluation of the response via Fourier analysis. However, the question whether a given response is statistically significant or not has received little attention so far, although it is especially relevant in high noise, low amplitude recordings, as often occur in pathologic conditions. A given response is statistically significant if it is unlikely that its value is due to noise fluctuations. Thus appropriate estimates of noise and response are required. We have analytically derived formulas for the statistical significance of a given signal-to-noise-ratio s, based on two different estimates of noise: (1) Noise estimate by a 'no stimulus' recording, or by a '+/- average'. The former needs an additional recording, the latter can simultaneously be calculated as the standard average. (2) Noise is estimated as the average of the two neighboring spectral lines (one below, and one above the response frequency). Analytical solutions were obtained for both noise estimates that can easily be evaluated in all appropriate recordings. Noise estimate (1) performs much poorer than noise estimate (2), as can be seen from the following landmark values: Typical significance levels of 5%, 1%, and 0.1% require s values of 4.36, 9.95, and 31.6 (1), and 2.82, 4.55, and 8.40 (2). The noise estimate based on the neighboring frequencies can be easily applied after recording, provided that the noise spectrum is reasonably smooth around the response and frequency-overspill was avoided. It allows a quantitative assessment of low responses in physiological threshold analyses and pathological conditions, e.g., 'submicrovolt flicker-ERG'.
引用
收藏
页码:207 / 232
页数:26
相关论文
共 23 条
[1]   A CLOSED-FORM APPROXIMATION TO ROBERTSON DETECTION CHARACTERISTICS [J].
ALBERHSEIM, WJ .
PROCEEDINGS OF THE IEEE, 1981, 69 (07) :839-839
[2]  
[Anonymous], 1989, HUMAN BRAIN ELECTROP
[3]   RETINAL AND CORTICAL ACTIVITY IN HUMAN-SUBJECTS DURING COLOR FLICKER FUSION [J].
BACH, M ;
GERLING, J .
VISION RESEARCH, 1992, 32 (07) :1219-1223
[4]  
Bach M., 1998, IOVS, V39, pS184
[5]  
Birch David G., 1996, Documenta Ophthalmologica, V92, P269, DOI 10.1007/BF02584081
[6]   Standard for visual evoked potentials 1995 [J].
Harding, GFA ;
Odom, JV ;
Spileers, W ;
Spekreijse, H .
VISION RESEARCH, 1996, 36 (21) :3567-3572
[7]   A NEW LASER INTERFEROMETER FOR THE STIMULATION OF PATTERN-REVERSAL VISUAL EVOKED-POTENTIALS [J].
KUSEL, R ;
WESEMANN, W ;
RASSOW, B .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1985, 6 (03) :239-246
[8]   Guidelines for basic pattern electroretinography - Recommendations by the International Society for Clinical Electrophysiology of Vision [J].
Marmor, MF ;
Holder, GE ;
Porciatti, V ;
Trick, GL ;
Zrenner, E .
DOCUMENTA OPHTHALMOLOGICA, 1995, 91 (04) :291-298
[9]   STANDARD FOR CLINICAL ELECTRORETINOGRAPHY (1994 UPDATE) [J].
MARMOR, MF ;
ZRENNER, E .
DOCUMENTA OPHTHALMOLOGICA, 1995, 89 (03) :199-210
[10]   FLUCTUATIONS OF STEADY-STATE VEPS - INTERACTION OF DRIVEN EVOKED-POTENTIALS AND THE EEG [J].
MAST, J ;
VICTOR, JD .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1991, 78 (05) :389-401