Linewidth measurement of free-running, pulsed, distributed feedback quantum cascade lasers

被引:11
作者
Beyer, T [1 ]
Braun, M [1 ]
Hartwig, S [1 ]
Lambrecht, A [1 ]
机构
[1] Fraunhofer Inst Phys Measurement Tech, D-79110 Freiburg, Germany
关键词
D O I
10.1063/1.1689750
中图分类号
O59 [应用物理学];
学科分类号
摘要
In gas spectroscopy with distributed feedback quantum cascade lasers (DFB-QCLs), linewidth is an important measurement parameter. However, in standard usage, these lasers are used in pulsed operation, so that the laser linewidth can no longer be assumed to remain constant over the length of the pulse. For the instantaneous measurement of such dynamic linewidths, a suitable measurement technique was still lacking. Frequently, pulse lengths of 5 to 10 ns are used to obtain laser linewidths that will be suitable for gas measurements. In these experiments, standard photoconductive or photovoltaic mid-infrared detectors with bandwidths in the 1 MHz range are employed, so that the linewidth is averaged over the tuning range of the QCL. The linewidth averaging could be reduced by shorter pulses, but for pulses shorter than 5 ns, the Fourier limitation increases the linewidth again. We present a technique to measure the dynamic laser linewidth continuously within a laser pulse of 30 ns to several mus. Due to the use of high-bandwidth detectors working from 2 to 12 mum, laser linewidths can be determined much more accurately than with short pulses. Instantaneous linewidths of DFB-QCLs down to 60 MHz are obtained. This shows that QCLs are well suited for gas measurement in the mid-infrared. (C) 2004 American Institute of Physics.
引用
收藏
页码:4551 / 4554
页数:4
相关论文
共 12 条
[1]   Gas detection with quantum cascade lasers: An adapted photoacoustic sensor based on Helmholtz resonance [J].
Barbieri, S ;
Pellaux, JP ;
Studemann, E ;
Rosset, D .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (06) :2458-2461
[2]   Fast gas spectroscopy using pulsed quantum cascade lasers [J].
Beyer, T ;
Braun, M ;
Lambrecht, A .
JOURNAL OF APPLIED PHYSICS, 2003, 93 (06) :3158-3160
[3]   Investigation of the spectral width of quantum cascade laser emission near 5.2 μm by a heterodyne experiment [J].
Ganser, H ;
Frech, B ;
Jentsch, A ;
Mürtz, M ;
Gmachl, C ;
Capasso, F ;
Sivco, DL ;
Baillargeon, JN ;
Hutchinson, AL ;
Cho, AY ;
Urban, W .
OPTICS COMMUNICATIONS, 2001, 197 (1-3) :127-130
[4]   Photoacoustic spectroscopy with quantum cascade distributed-feedback lasers [J].
Hofstetter, D ;
Beck, M ;
Faist, J ;
Nägele, M ;
Sigrist, MW .
OPTICS LETTERS, 2001, 26 (12) :887-889
[5]   Methane concentration and isotopic composition measurements with a mid-infrared quantum-cascade laser [J].
Kosterev, AA ;
Curl, RF ;
Tittel, FK ;
Gmachl, C ;
Capasso, F ;
Sivco, DL ;
Baillargeon, JN ;
Hutchinson, AL ;
Cho, AY .
OPTICS LETTERS, 1999, 24 (23) :1762-1764
[6]  
Kosterev AA, 2001, LASER PHYS, V11, P39
[7]   Cavity ringdown spectroscopic detection of nitric oxide with a continuous-wave quantum-cascade laser [J].
Kosterev, AA ;
Malinovsky, AL ;
Tittel, FK ;
Gmachl, C ;
Capasso, F ;
Sivco, DL ;
Baillargeon, JN ;
Hutchinson, AL ;
Cho, AY .
APPLIED OPTICS, 2001, 40 (30) :5522-5529
[8]   Free-running frequency stability of mid-infrared quantum cascade lasers [J].
Myers, TL ;
Williams, RM ;
Taubman, MS ;
Gmachl, C ;
Capasso, F ;
Sivco, DL ;
Baillargeon, JN ;
Cho, AY .
OPTICS LETTERS, 2002, 27 (03) :170-172
[9]   Sensitive absorption spectroscopy with a room-temperature distributed-feedback quantum-cascade laser [J].
Namjou, K ;
Cai, S ;
Whittaker, EA ;
Faist, J ;
Gmachl, C ;
Capasso, F ;
Sivco, DL ;
Cho, AY .
OPTICS LETTERS, 1998, 23 (03) :219-221
[10]   Sub-part-per-billion detection of nitric oxide in air using a thermoelectrically cooled mid-infrared quantum cascade laser spectrometer [J].
Nelson, DD ;
Shorter, JH ;
McManus, JB ;
Zahniser, MS .
APPLIED PHYSICS B-LASERS AND OPTICS, 2002, 75 (2-3) :343-350