Evaluation of a dual-arm Archimedean spiral array for microwave hyperthermia

被引:35
作者
Johnson, J. E.
Neuman, D. G.
Maccarini, P. F.
Juang, T.
Stauffer, P. R.
Turner, P.
机构
[1] Univ Calif San Francisco, San Francisco, CA 94143 USA
[2] BSD Med Corp, Salt Lake City, UT USA
关键词
superficial hyperthermia; microwave antenna arrays; multi-element array applicator; spiral microstrip antennae;
D O I
10.1080/02656730600905595
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: This effort describes a third-party performance evaluation of a novel, commercial, dual-armed Archimedean spiral array hyperthermia applicator. The applicator is analysed for its ability to couple efficiently into muscle equivalent phantom loads, operate over a broad bandwidth to help accommodate variable tissue properties and generate predictable and repeatable SAR contours that are adaptable to clinically probable disease shapes. Materials and methods: Characterization of the applicator includes E-field and return-loss measurements in liquid muscle tissue-equivalent phantom, as well as comparison of 'treatment-planning' simulations of several possible array SAR patterns with measured SAR from non-coherently driven spiral array antennae. Results: The applicator demonstrates a reasonably low return loss over a large bandwidth and the ability to generate a very uniform heating pattern. Ability to adjust SAR contours spatially to fit specific shapes is also demonstrated. Conclusions: This device should prove a welcome addition to a currently limited set of superficial heating applicators to provide controllable heating of superficial tissue disease.
引用
收藏
页码:475 / 490
页数:16
相关论文
共 41 条
[1]  
*BSD, 2005, SA812 BSD MED CORP
[2]   FDTD SIMULATIONS OF CLINI-THERM APPLICATORS ON INHOMOGENEOUS PLANAR TISSUE MODELS [J].
CHAN, KW ;
MCDOUGALL, JA ;
CHOU, CK .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1995, 11 (06) :809-820
[3]   EVALUATION OF CAPTIVE BOLUS APPLICATORS [J].
CHOU, CK ;
MCDOUGALL, JA ;
CHAN, KW ;
LUK, KH .
MEDICAL PHYSICS, 1990, 17 (04) :705-709
[4]   Effects of waterbolus size, shape and configuration on the SAR distribution pattern of the Lucite cone applicator [J].
de Bruijne, M ;
Samaras, T ;
Bakker, JF ;
Van Rhoon, GC .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2006, 22 (01) :15-28
[5]   PRECLINICAL EVALUATION OF A MICROWAVE PLANAR ARRAY APPLICATOR FOR SUPERFICIAL HYPERTHERMIA [J].
DIEDERICH, CJ ;
STAUFFER, PR .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1993, 9 (02) :227-246
[6]  
ERB J, 1995, SIMULATION KONSTRUKT
[7]   An attempt at quantitative specification of SAR distribution homogeneity [J].
Gelvich, EA ;
Mazokhin, VN ;
Troshin, II .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1996, 12 (03) :431-436
[8]   Contact flexible microstrip applicators (CFMA) in a range from microwaves up to short waves [J].
Gelvich, EA ;
Mazokhin, VN .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (09) :1015-1023
[9]   Resonance effects in applicator water boluses and their influence on SAR distribution patterns [J].
Gelvich, EA ;
Mazokhin, VN .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2000, 16 (02) :113-128
[10]   CURRENT SHEET APPLICATORS FOR CLINICAL MICROWAVE HYPERTHERMIA [J].
GOPAL, MK ;
CETAS, TC .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1993, 41 (03) :431-437