Accuracy of the patient-model is a critical point in robot assisted surgery. When performing craniotomies, the dura mater must not be perforated. Hence bone width is of particular interest. The influence of imaging and segmentation on accuracy of the width of the bone-model was investigated. A human cadaver head was scanned with a CT-scanner under a variety of image acquisition parameters. Bone was segmented from these image data sets using threshold based segmentation with different settings for the lower threshold. From these volume data sets surface models of the bone were generated. The real width of the bone of the skull was measured at several positions. Using fiducial marker registration, these measured values were compared to the corresponding positions in the bone-models. CT-scan imaging with a slice thickness and slice distance of 1.5 to 2mm and a segmentation of bone with a lower threshold of 300 or 400 Hounsfield Units resulted in models with an average accuracy of 0.4mm for bone-width. However, at some points these models were too thin by up to 0.9mm. More accurate models are needed. It has to be evaluated, whether CT imaging with higher resolution or more sophisticated segmentation algorithms can reduce the scatter.
引用
收藏
页码:829 / 836
页数:8
相关论文
共 13 条
[1]
Determination of the Hounsfield value for CT-based design of custom femoral stems [J].
Aamodt, A ;
Kvistad, KA ;
Andersen, E ;
Lund-Larsen, J ;
Eine, J ;
Benum, P ;
Husby, OS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (01) :143-147
[2]
In vivo measurements of temporal bone on reconstructed clinical high-resolution computed tomography scans [J].
Chan, LL ;
Manolidis, S ;
Taber, KH ;
Hayman, LA .
LARYNGOSCOPE, 2000, 110 (08) :1375-1378
[3]
Computer-aided surgical planning for implantation of hearing aids based on CT data in a VR environments [J].
Dammann, F ;
Bode, A ;
Schwaderer, E ;
Schaich, M ;
Heuschmid, M ;
Maassen, MM .
RADIOGRAPHICS, 2001, 21 (01) :183-190
[4]
Dauber S, 2002, Biomed Tech (Berl), V47 Suppl 1 Pt 2, P936, DOI 10.1515/bmte.2002.47.s1b.936
[5]
Engel D, 2001, IEEE INT CONF ROBOT, P2020, DOI 10.1109/ROBOT.2001.932904
[6]
Predicting error in rigid-body point-based registration [J].
Fitzpatrick, JM ;
West, JB ;
Maurer, CR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) :694-702
[7]
CT number variations due to different image acquisition and reconstruction parameters: a thorax phantom study [J].
Groell, R ;
Rienmueller, R ;
Schaffler, GJ ;
Portugaller, HR ;
Graif, E ;
Willfurth, P .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2000, 24 (02) :53-58
[8]
Hothan T, 2001, RADIOLOGE, V41, P497, DOI 10.1007/s001170051061
[9]
Accuracy of object depiction and opacity transfer function optimization in CT volume-rendered images [J].
Jani, AB ;
Pelizzari, CA ;
Chen, GTY ;
Grzeszczuk, RP .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1998, 22 (03) :459-470
[10]
KALENDER W, 2002, COMPUTERTOMOGRAPHIE, P30