Dual-anode high-pressure xenon cylindrical ionization chamber

被引:13
作者
Bolotnikov, A [1 ]
Bolozdynya, A
DeVito, R
Richards, J
机构
[1] Brookhaven Natl Lab, Upton, NY 11973 USA
[2] Constellat Technol Corp, Largo, FL 33777 USA
关键词
Frisch-grid; gamma rays; ionization chamber; spectrometer; Xenon;
D O I
10.1109/TNS.2004.829369
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A new approach to the design of high-pressure xenon cylindrical ionization chambers is investigated. In the new approach, a dual anode is used to replace the single anode surrounded by a shielding-grid, which are employed in a conventional design. Two anode wires are stretched near the axis of the 30 cm long cylindrical chamber with a 9 cm diameter cathode. Both the wires are kept at the same (nearly zero) potential and are dc-coupled to the charge-sensitive preamplifiers. For most cases, only one of the wires (it can be either one) collects the electrons produced by. an ionizing event. The difference between the signals read out from the wires is proportional to the total produced charge. The experimental results are compared to Monte Carlo simulations. The optimal design of the dual-anode cylindrical ionization chamber is discussed.
引用
收藏
页码:1262 / 1269
页数:8
相关论文
共 17 条
[1]   Improving the energy resolution of high-pressure Xe cylindrical ionization chambers [J].
Bolotnikov, A ;
Ramsey, B .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1006-1010
[2]  
BOLOTNIKOV AE, 1986, PRIB TEKH EKSP, V29, P42
[3]  
BOLOTNIKOV AE, 1992, P INT C LIQ RAD DET, P462
[4]   A high pressure xenon self-triggered scintillation drift chamber with 3D sensitivity in the range of 20-140 keV deposited energy [J].
Bolozdynya, A ;
Egorov, V ;
Koutchenkov, A ;
Safronov, G ;
Smirnov, G ;
Medved, S ;
Morgunov, V .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 385 (02) :225-238
[5]  
BOLOZDYNYA A, 2002, P INMM 43 ANN M ORL, P102
[6]  
Dmitrenko V. V., 1983, Soviet Physics - Technical Physics, V28, P1440
[7]  
GALPER AM, 1981, P 17 ICRC C LA JOLL, P287
[8]   MONTE-CARLO SIMULATION OF ELECTRON TRAJECTORIES IN HIGH-PRESSURE XENON GAS [J].
IQBAL, MZ ;
OCALLAGHAN, BMG ;
WONG, HT .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 253 (02) :278-287
[9]   A hemispherical high-pressure xenon gamma radiation spectrometer [J].
Kessick, R ;
Tepper, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 490 (1-2) :243-250
[10]  
KNOLL GF, 2000, RAD DETECTION MEASUR, P148