Larger bone graft size and washing of bone grafts prior to impaction enhances the initial stability of cemented cups - Experiments using a synthetic acetabular model

被引:51
作者
Arts, J. J. Chris [1 ]
Verdonschot, Nico [1 ]
Buma, Pieter [1 ]
Schreurs, B. Willem [1 ]
机构
[1] Radboud Univ Nijmegen, Nijmegen Med Ctr, Orthopaed Res Lab, NL-6500 HB Nijmegen, Netherlands
关键词
D O I
10.1080/17453670610045957
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Bone defects after failed total hip arthroplasty can be reconstructed with impacted morselized bone grafts and a cemented cup. On the acetabular side, the effects on initial cup stability of washing bone grafts prior to impaction and bone graft size remain unclear. Related to these variables, cement penetration and inter-particle shear resistance have been suggested to be critical factors to ensure initial cup stability. Methods Using a synthetic acetabular model, we studied the effects of (1) washing bone grafts prior to impaction, and (2) bone graft size on the initial stability of cemented cups. In addition, cement penetration was measured using CT scans. Results Reconstructions with large, washed bone grafts provided the highest stability during mechanical compression and in a lever-out situation. Washing of the bone grafts had a positive effect on initial cup stability, but the size of the bone grafts appeared to be more important. Cement penetration was affected by bone graft size but not by washing. Interpretation From a mechanical standpoint, large bone grafts that have been washed prior to impaction may be preferable in order to obtain optimal cup stability using the bone impaction grafting technique.
引用
收藏
页码:227 / 233
页数:7
相关论文
共 18 条
[1]   Acetabular defect reconstruction with impacted morsellized bone grafts or TCP/HA particles. A study on the mechanical stability of cemented cups in an artificial acetabulum model [J].
Bolder, SBT ;
Verdonschot, N ;
Schreurs, BW ;
Buma, P .
BIOMATERIALS, 2002, 23 (03) :659-666
[2]   Particle size of bone graft and method of impaction affect initial stability of cemented cups - Human cadaveric and synthetic pelvic specimen studies [J].
Bolder, SBT ;
Schreurs, BW ;
Verdonschot, N ;
van Unen, JMJ ;
Gardeniers, JWM ;
Slooff, TJJH .
ACTA ORTHOPAEDICA SCANDINAVICA, 2003, 74 (06) :652-657
[3]   Mechanical considerations in impaction bone grafting [J].
Brewster, NT ;
Brewster, WJ ;
Howie, CR ;
Madabhushi, SPG ;
Usmani, AS ;
Fairbairn, DR .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (01) :118-124
[4]   Mechanical behavior of human morselized cancellous bone in triaxial compression testing [J].
Brodt, MD ;
Swan, CC ;
Brown, TD .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1998, 16 (01) :43-49
[5]  
CROWNINSHIELD RD, 1998, CLIN ORTHOP RELAT R, V355, P90
[6]   Techniques to improve the shear strength of impacted bone graft - The effect of particle size and washing of the graft [J].
Dunlop, DG ;
Brewster, NT ;
Madabhushi, SPG ;
Usmani, AS ;
Pankaj, P ;
Howie, CR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (04) :639-646
[7]  
GROSS AE, 1999, CLIN ORTHOP RELAT R, V369, P198
[8]   High-pressure saline washing of allografts reduces bacterial contamination [J].
Hirn, MYJ ;
Salmela, PM ;
Vuento, RE .
ACTA ORTHOPAEDICA SCANDINAVICA, 2001, 72 (01) :83-85
[9]   Impaction technique and graft treatment in revisions of the femoral component -: Laboratory studies and clinical validation [J].
Höstner, J ;
Hultmark, P ;
Kärrholm, J ;
Malchau, H ;
Tveit, M .
JOURNAL OF ARTHROPLASTY, 2001, 16 (01) :76-82
[10]  
Ornstein E, 2002, ACTA ORTHOP SCAND, V73, P3