Endovascular coil occlusion of intracranial aneurysms assisted by a novel self-expandable nitinol microstent (Neuroform)

被引:115
作者
Henkes, H
Bose, A
Felber, S
Miloslavski, E
Berg-Dammer, E
Kühne, D
机构
[1] Alfried Krupp Hosp, Klin Radiol & Neuroradiol, D-45117 Essen, Germany
[2] Alfried Krupp Hosp, Neurol Klin Neurophysiol, D-45117 Essen, Germany
[3] NYU Med Ctr, Neurointervent Serv, New York, NY 10016 USA
[4] Smart Therapeut, San Leandro, CA USA
来源
INTERVENTIONAL NEURORADIOLOGY | 2002年 / 8卷 / 02期
关键词
intracranial aneurysm; electrolytically detachable coil; intracranial stent; nitinol; embolization; internal carotid artery; vertebral artery; basilar artery;
D O I
10.1177/159101990200800202
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Endovascular treatment of wide neck intracranial aneurysms is technically difficult and leads to less favorable treatment results and long term outcome. We participated in a multicenter prospective study to evaluate the safety and performance of a new self-expandable nitinol micro stent (Neuroform) in stent assisted coil occlusion of wide neck intracranial aneurysms. Eighteen patients were enrolled in the study in a single center The anatomy of the target aneurysm and the parent vessel, technical details of the procedure, device functionality, anatomic and clinical results were evaluated. All enrolled aneurysms were either wide necked or showed an unfavorable neck-to-fundus ratio. In 16 out of 18 patients the Neuroform device allowed stent assisted coil occlusion of the aneurysm. The occlusion rate was 95% in eight patients and 100% in eight patients. The two failures were both due to anatomic reasons. Flexibility of the stent, behavior during deployment and subsequent ability to retain coils within the aneurysmal sac were considered as good as or better than the properties of previous balloon expandable stents. No device-related adverse events were encountered. Procedure-related clinical complications occurred in seven patients but caused no severe permanent neurological deficit. The Neuroform neurovascular stenting system is a safe and effective adjunct for the stent-assisted coil occlusion of wide necked intracranial aneurysms. The major advantages of this device are its self-expanding property and very high flexibility which allows safe navigation, easy sizing, as well as accurate positioning of the stent while providing sufficient bridging of the aneurysm neck for subsequent coil placement.
引用
收藏
页码:107 / 119
页数:13
相关论文
共 36 条
  • [1] Endovascular repair of traumatic pseudoaneurysm by uncovered self-expandable stenting with or without transstent coiling of the aneurysm cavity
    Assali, AR
    Sdringola, S
    Moustapha, A
    Rihner, M
    Denktas, AE
    Lefkowitz, MA
    Campbell, M
    Smalling, RW
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2001, 53 (02) : 253 - 258
  • [2] NEOINTIMA FORMATION AFTER VASCULAR STENT IMPLANTATION - SPATIAL AND CHRONOLOGICAL DISTRIBUTION OF SMOOTH-MUSCLE CELL-PROLIFERATION AND PHENOTYPIC MODULATION
    BAI, HZ
    MASUDA, J
    SAWA, Y
    NAKANO, S
    SHIRAKURA, R
    SHIMAZAKI, Y
    OGATA, J
    MATSUDA, H
    [J]. ARTERIOSCLEROSIS AND THROMBOSIS, 1994, 14 (11): : 1846 - 1853
  • [3] Cloft HJ, 2000, AM J NEURORADIOL, V21, P1312
  • [4] Intracranial stent placement to trap an extruded coil during endovascular aneurysm treatment: Technical note
    Fessler, RD
    Ringer, AJ
    Qureshi, AI
    Guterman, LR
    Hopkins, LN
    [J]. NEUROSURGERY, 2000, 46 (01) : 248 - 251
  • [5] GEREMIA G, 1994, AM J NEURORADIOL, V15, P1223
  • [6] Elective stenting of symptomatic basilar artery stenosis
    Gomez, CR
    Misra, VK
    Liu, MW
    Wadlington, VR
    Terry, JB
    Tulyapronchote, R
    Campbell, MS
    [J]. STROKE, 2000, 31 (01) : 95 - 99
  • [7] Natural history of the neck remnant of a cerebral aneurysm treated with the Guglielmi detachable coil system
    Hayakawa, M
    Murayama, Y
    Duckwiler, GR
    Gobin, YP
    Guglielmi, G
    Viñuela, F
    [J]. JOURNAL OF NEUROSURGERY, 2000, 93 (04) : 561 - 568
  • [8] Intravascular stent and endovascular coil placement for a ruptured fusiform aneurysm of the basilar artery - Case report and review of the literature
    Higashida, RT
    Smith, W
    Gress, D
    Urwin, R
    Dowd, CF
    Balousek, PA
    Halbach, VV
    [J]. JOURNAL OF NEUROSURGERY, 1997, 87 (06) : 944 - 949
  • [9] Transluminal stent-assisted coil embolization of a vertebral confluence aneurysm: Technique report
    Horowitz, MB
    Levy, EI
    Koebbe, CJ
    Jungreis, CC
    [J]. SURGICAL NEUROLOGY, 2001, 55 (05): : 291 - 296
  • [10] Efficacy and current limitations of intravascular stents for intracranial internal carotid, vertebral, and basilar artery aneurysms
    Lanzino, G
    Wakhloo, AK
    Fessler, RD
    Hartney, ML
    Guterman, LR
    Hopkins, LN
    [J]. JOURNAL OF NEUROSURGERY, 1999, 91 (04) : 538 - 546