Noninvasive determination of cardiac output using single breath CO2 analysis

被引:11
作者
Arnold, JH
Stenz, RI
Thompson, JE
Arnold, LW
机构
[1] CHILDRENS HOSP, DEPT ANESTHESIA, MULTIDISCIPLINARY INTENS CARE UNIT, FARLEY 5, 300 LONGWOOD AVE, BOSTON, MA 02115 USA
[2] CHILDRENS HOSP, DEPT CARDIOL, BOSTON, MA 02115 USA
[3] CHILDRENS HOSP, DEPT RESP CARE, BOSTON, MA 02115 USA
[4] HARVARD UNIV, SCH MED, DEPT ANESTHESIA, BOSTON, MA 02115 USA
[5] HARVARD UNIV, SCH MED, DEPT PEDIAT, BOSTON, MA 02115 USA
关键词
carbon dioxide; cardiac output; respiratory deadspace; gas exchange; mechanical ventilation; respiration disorders; ventilation perfusion ratio; lungs; critical illness;
D O I
10.1097/00003246-199610000-00017
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To examine the utility of single breath CO2 analysis as a noninvasive measure of cardiac output. Setting: An animal laboratory in a university-affiliate medical center. Design: A prospective, animal cohort study comparing 21 parameters derived from single breath CO2 analysis with cardiac output determined by an ultrasonic flow probe. Subjects: Six healthy adult sheep. Methods: The single breath CO2 analysis station consists of a mainstream capnometer, a variable orifice pneumotachometer, a signal processor, and computer software with capability for both on- and off line data analysis, Twenty-one derived components of the CO2 expirogram were evaluated as predictors of cardiac output, Cardiac output was manipulated by successive injections of a hydraulic constrictor placed around the inferior vena cava. Measurements and Main Results: Thirty four measurements of cardiac output were available for comparison with derived variables from the CO2 expirogam, Stepwise linear regression identified two variables that were most predictive of cardiac output: a) the angle between the slope lines for phases II and III of the CO2 expirogram divided by the volume of CO2 per breath (angle/mL CO2); and b) the slope of phase II, The multivariate equation was highly statistically significant and explained 94% of the variance (adjusted r(2) = .94, p < .0001). The bias and precision of the calculated cardiac output were .00 and .23, respectively, The mean percent difference for the cardiac output estimate derived from the single breath CO2 analysis station was 0.36%. Conclusions: Our data indicate that analysis of the CO2 expirogram can yield accurate information about the cardiovascular system, Specifically, two variables derived from a plot of expired CO2 concentration vs, expired volume predict changes in cardiac output in healthy adult sheep with an adjusted coefficient of determination of .94. Prospective application of this technology in the setting of lung injury and rapidly changing physiology will be essential in determining the clinical usefulness of the technique.
引用
收藏
页码:1701 / 1705
页数:5
相关论文
共 25 条
[1]   On the fluctuation in the composition of the alveolar air during the respiratory cycle in muscular exercise. [J].
Aitken, RS ;
Clark-Kennedy, AE .
JOURNAL OF PHYSIOLOGY-LONDON, 1928, 65 (04) :389-411
[2]   Single breath CO2 analysis: Description and validation of a method [J].
Arnold, JH ;
Thompson, JE ;
Arnold, LW .
CRITICAL CARE MEDICINE, 1996, 24 (01) :96-102
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   NONINVASIVE MEASUREMENT OF CARDIAC-OUTPUT USING PARTIAL CO2 REBREATHING [J].
CAPEK, JM ;
ROY, RJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (09) :653-661
[5]   CARDIAC OUTPUT IN MIDDLE-AGED PATIENTS DETERMINED WITH CO2 REBREATHING METHOD [J].
CLAUSEN, JP ;
LARSEN, OA ;
TRAPJENS.J .
JOURNAL OF APPLIED PHYSIOLOGY, 1970, 28 (03) :337-+
[6]   DETERMINATION OF MIXED VENOUS CO2 TENSIONS BY REBREATHING [J].
COLLIER, CR .
JOURNAL OF APPLIED PHYSIOLOGY, 1956, 9 (01) :25-29
[7]   MEASUREMENTS OF CARDIAC-OUTPUT IN SERIOUSLY ILL PATIENTS USING A CO2 REBREATHING METHOD [J].
DAVIS, CC ;
JONES, NL ;
SEALEY, BJ .
CHEST, 1978, 73 (02) :167-172
[8]   THE CONCEPT OF DEADSPACE WITH SPECIAL REFERENCE TO THE SINGLE BREATH TEST FOR CARBON-DIOXIDE [J].
FLETCHER, R ;
JONSON, B ;
CUMMING, G ;
BREW, J .
BRITISH JOURNAL OF ANAESTHESIA, 1981, 53 (01) :77-88
[9]  
FLETCHER R, 1986, ANESTH ANALG, V65, P645
[10]   DEADSPACE AND THE SINGLE BREATH TEST FOR CARBON-DIOXIDE DURING ANESTHESIA AND ARTIFICIAL-VENTILATION [J].
FLETCHER, R ;
JONSON, B .
BRITISH JOURNAL OF ANAESTHESIA, 1984, 56 (02) :109-119