Eliminating amino acid and peptide interference in high-performance anion-exchange pulsed amperometric detection glycoprotein monosaccharide analysis

被引:29
作者
Weitzhandler, M
Pohl, C
Rohrer, J
Narayanan, L
Slingsby, R
Avdalovic, N
机构
[1] Dionex Corporation, Sunnyvale, CA 94086
关键词
D O I
10.1006/abio.1996.0386
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The monosaccharide content of a glycoprotein is often determined by acid hydrolysis at elevated temperature and subsequent high pH chromatography of the released, underivatized monosaccharides on pellicular anion-exchange resin (HPAE) using pulsed amperometric detection (PAD). We have found that for glycoproteins with low levels of glycosylation, monosaccharide quantitation can be compromised by amino acids fouling the working electrode surface. Specifically, lysine elutes on the CarboPac PA1 column just prior to galactosamine, whereas remaining amino acids and most peptides elute after the monosaccharides and do not interfere with monosaccharide quantification. A direct comparison of PAD vs Abs(215) detection of lysine using the CarboPac PA1 column as the separator reveals that lysine does not cleanly come off the working electrode. The monosaccharide response inhibition caused by lysine could be corrected by the posthydrolysis addition of a rhamnose internal standard and the determination of ''correction factors.'' We have developed a guard column with an altered selectivity for amino acids which, when used with a new separator, causes lysine to elute after the monosaccharides and also causes hydrophobic amino acids to elute further after the monosaccharides. Together the new separator and guard columns solve the lysine fouling problem, reduce sample-related baseline noise, and reduce the magnitude of correction factors. (C) 1996 Academic Press, Inc.
引用
收藏
页码:128 / 134
页数:7
相关论文
共 7 条
[1]   QUANTITATIVE-DETERMINATION OF MONOSACCHARIDES IN GLYCOPROTEINS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH HIGHLY SENSITIVE FLUORESCENCE DETECTION [J].
ANUMULA, KR .
ANALYTICAL BIOCHEMISTRY, 1994, 220 (02) :275-283
[2]  
*DION, 1994, 21 DION CORP
[3]   PULSED ELECTROCHEMICAL DETECTION AT THE RING OF A RING-DISK ELECTRODE APPLIED TO A STUDY OF AMINE ADSORPTION AT GOLD ELECTRODES [J].
DOBBERPUHL, DA ;
JOHNSON, DC .
ANALYTICAL CHEMISTRY, 1995, 67 (07) :1254-1258
[4]   MONOSACCHARIDE ANALYSIS OF GLYCOCONJUGATES BY ANION-EXCHANGE CHROMATOGRAPHY WITH PULSED AMPEROMETRIC DETECTION [J].
HARDY, MR ;
TOWNSEND, RR ;
LEE, YC .
ANALYTICAL BIOCHEMISTRY, 1988, 170 (01) :54-62
[5]  
JOHNSON DC, 1994, CARBOHYDRATE ANAL HI
[6]  
ROHRER J, 1995, TECH PROT CHEM, V6, P65, DOI 10.1016/S1080-8914(06)80011-6
[7]  
SINGLETON JA, 1993, 206 AM CHEM SOC NAT