Electromigration dispersion in capillary zone electrophoresis -: Experimental validation of use of the Haarhoff-Van der Linde function

被引:43
作者
Erny, GL [1 ]
Bergström, ET [1 ]
Goodall, DM [1 ]
机构
[1] Univ York, Dept Chem, York YO10 5DD, N Yorkshire, England
关键词
dispersion; validation; Haarhoff-Van der Linde function; peak shape; paraquat; aminopyridines; pesticides; pyridines;
D O I
10.1016/S0021-9673(02)00454-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This paper provides experimental validation of the use of the Haarhoff-Van der Linde (HVL) peak fitting function to fit experimental capillary zone electrophoresis (CZE) electropherogrants. The test mixtures were composed of paraquat over a five order of magnitude concentration range (1.2 muM to 120 mM) and 4-aminopyridine at constant concentration (0.53 mM) as internal standard. Peak descriptors and electrophoresis parameters were extracted reliably by a Gaussian function from 4 to 40 muM; by the HVL function from 120 muM to 4 mM; and by a triangular function from 4 to 120 mM. The HVL function can be used where there is significant peak asymmetry due to electromigration distortion (EMD) and the Gaussian contribution toward the peak variance is greater than 25%. The peak centre (a(1)) and the Gaussian variance (a(2)) of the paraquat peak are shown to be independent of concentration. Diffusion coefficients obtained from a(2) for both analytes were found to be in good agreement with their theoretical values. For all peaks where the distortion coefficient (a(3)) can be extracted, this parameter is shown to be directly proportional to the sample loading, as predicted by EMD theory. For the 4-aminopyridinium ion, mobilities calculated from a(3) and measured independently are in excellent agreement. These results show that the HVL function accurately describes the two major processes, diffusion and EMD, contributing to the variance during a CZE separation. (C) 2002 Published by Elsevier Science B.V.
引用
收藏
页码:229 / 239
页数:11
相关论文
共 19 条
[1]   A charge coupled device array detector for single-wavelength and multiwavelength ultraviolet absorbance in capillary electrophoresis [J].
Bergstrom, ET ;
Goodall, DM ;
Pokric, B ;
Allinson, NM .
ANALYTICAL CHEMISTRY, 1999, 71 (19) :4376-4384
[2]  
BERGSTROM ET, 2002, IN PRESS ANAL CHEM
[3]   Salt effects in capillary zone electrophoresis II. Mechanisms of electrophoretic mobility modification due to Joule heating at high buffer concentrations [J].
Cross, RF ;
Cao, J .
JOURNAL OF CHROMATOGRAPHY A, 1998, 809 (1-2) :159-171
[4]   Predicting peak shape in capillary zone electrophoresis:: a generic approach to parametrizing peaks using the Haarhoff-Van der Linde (HVL) function [J].
Erny, GL ;
Bergström, ET ;
Goodall, DM .
ANALYTICAL CHEMISTRY, 2001, 73 (20) :4862-4872
[5]   EQUATIONS FOR CALCULATION OF CHROMATOGRAPHIC FIGURES OF MERIT FOR IDEAL AND SKEWED PEAKS [J].
FOLEY, JP ;
DORSEY, JG .
ANALYTICAL CHEMISTRY, 1983, 55 (04) :730-737
[6]   CONCENTRATION DEPENDENCE OF ELUTION CURVES IN NON-IDEAL GAS CHROMATOGRAPHY [J].
HAARHOFF, PC ;
VANDERLI.HJ .
ANALYTICAL CHEMISTRY, 1966, 38 (04) :573-&
[7]  
Kenndler E, 1998, CHEM ANAL SERIES MON, V146, P25
[8]  
Klinkenberg A., 1960, GAS CHROMATOGRAPHY
[9]   THERMAL EFFECTS AND BAND SPREADING IN CAPILLARY ELECTRO-SEPARATION [J].
KNOX, JH .
CHROMATOGRAPHIA, 1988, 26 :329-337
[10]   Prediction of electrophoretic mobilities. 3. Effect of ionic strength in capillary zone electrophoresis [J].
Li, DM ;
Fu, SL ;
Lucy, CA .
ANALYTICAL CHEMISTRY, 1999, 71 (03) :687-699