Antibiotic susceptibility of glutaraldehyde-tolerant Mycobacterium chelonae from bronchoscope washing machines

被引:24
作者
Nomura, K
Ogawa, M
Miyamoto, H
Muratani, T
Taniguchi, H
机构
[1] Univ Occupat & Environm Hlth, Sch Med, Dept Microbiol, Yahatanishi Ku, Kitakyushu, Fukuoka 8077555, Japan
[2] Univ Occupat & Environm Hlth, Dept Urol, Kitakyushu, Fukuoka 8077555, Japan
关键词
D O I
10.1016/j.ajic.2003.07.007
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: Contamination of bronchoalveolar lavage fluid is a major problem in the world. Although 2% glutaraldehyde (GA) is widely used as a disinfectant for bronchoscope cleaning, recently, GA-tolerant mycobacteria have been isolated, which makes this problem more complicated. Methods: We studied the susceptibility to GA and antibiotics of mycobacteria isolated from bronchoscope washing machines in our hospital. We also studied the minimum inhibitory concentrations of GA and antibiotics with pump inhibitors. Results: Twenty-nine mycobacteria were isolated, of which 26 were Mycobacterium chelonae. Among 18 isolates of M chelonae, excluding 8 isolates in which some results were not reproducible, 50% (9 of 18) were 2% GA-tolerant. One hundred percent (9 of 9) of the GA-tolerant isolates and 11% (1 of 9) of the GA-sensitive isolates were either resistant or intermediately resistant to 2 or 3 classes of antibiotics. Efflux pump inhibitors did not influence the susceptibility to GA and antibiotics. Conclusions: It was suggested that there might be an association of GA tolerance with antibiotic resistance in M chelonae. There may a different mechanism(s) other than that involving efflux pumps with regard to GA tolerance and antibiotic resistance in M chelonae. When bronchoscopy-related mycobacterial infections are suspected, physicians and clinical microbiologists should exercise care in handling GA-tolerant mycobacteria, which may be resistant to multiple antibiotics.
引用
收藏
页码:185 / 188
页数:4
相关论文
共 15 条
  • [1] BODDINGHAUS B, 1990, J CLIN MICROBIOL, V28, P1751
  • [2] Covert TC, 1999, APPL ENVIRON MICROB, V65, P2492
  • [3] Cox R, 1997, INFECT CONT HOSP EP, V18, P136
  • [4] George AM, 1996, FEMS MICROBIOL LETT, V139, P1, DOI 10.1016/0378-1097(96)00127-9
  • [5] PSEUDOEPIDEMIC OF NONTUBERCULOUS MYCOBACTERIA DUE TO A CONTAMINATED BRONCHOSCOPE CLEANING MACHINE - REPORT OF AN OUTBREAK AND REVIEW OF THE LITERATURE
    GUBLER, JGH
    SALFINGER, M
    VONGRAEVENITZ, A
    [J]. CHEST, 1992, 101 (05) : 1245 - 1249
  • [6] HARANO Y, 1991, KOKKYU, V10, P320
  • [7] KLINGEREN BV, 1993, J HOSP INFECT, V25, P147
  • [8] KOGA T, 1986, J JAP SOC BRONCHOL, V8, P338
  • [9] KOGA T, 2000, J JPN SOC BRONCHOL, V22, P251
  • [10] Reduced glutaraldehyde susceptibility in Mycobacterium chelonae associated with altered cell wall polysaccharides
    Manzoor, SE
    Lambert, PA
    Griffiths, PA
    Gill, MJ
    Fraise, AP
    [J]. JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 1999, 43 (06) : 759 - 765