Effects of converting bilateral cochlear implant subjects to a strategy with increased rate and number of channels

被引:13
作者
Dunn, CC [1 ]
Tyler, RS [1 ]
Witt, SA [1 ]
Gantz, BJ [1 ]
机构
[1] Univ Iowa, Dept Otorhinolaryngol Head & Neck Surg, Iowa City, IA 52242 USA
关键词
bilateral cochlear implants; HiResolution strategy; sound processing; speech perception in noise;
D O I
10.1177/000348940611500605
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objectives: Three different Advanced Bionics processing strategies were evaluated: 1) 8-channel, 813 pulses per second (pps), Continuous Interleaved Sampling (CIS); 2) 16-channel, 5,100 pps, HiResolution Paired (HiRes P); and 3) 16-channel, 2,900 pps, HiResolution Sequential (HiRes S). Methods: Seven adult bilateral Clarion CII cochlear implant recipients who had been using a CIS processing strategy for at least 18 months participated in this study. Sentence recognition in multitalker babble from the front was collected on subjects using their CIS strategy and after subjects were programmed for the first time with HiRes P and HiRes S strategies. An ABAB design was implemented for 1 month whereby subjects used each HiResolution strategy every other day. Sentence recognition testing was repeated at the 1- and 3-month post-HiResolution test intervals. Results: Comparisons between the CIS and HiResolution strategies showed immediate improvements for 5 subjects in favor of the HiResolution strategies. After 1 month of alternating between the HiRes P and HiRes S strategies, remarkably, 2 subjects showed improvements of 60%, 2 subjects showed improvements of 40%, and 2 subjects showed improvements of 30% over the CIS strategy that they had previously used for at least 18 months. The results after 3 months of use were consistent with those obtained at 1 month. Conclusions: The HiRes S and HiRes P strategies resulted in dramatic improvements in speech perception in noise for a subset of subjects who had been using the CIS strategy bilaterally. This finding demonstrates that these subjects were able to tolerate a more difficult signal-to-noise ratio. Further work is needed to determine the independent effects of rate versus number of channels.
引用
收藏
页码:425 / 432
页数:8
相关论文
共 7 条
[1]  
[Anonymous], 1983, Iowa cochlear implant tests
[2]  
[Anonymous], 1985, SENTENCE TEST SPEECH
[3]   Speech perception results in children implanted with Clarion® devices:: Hi-Resolution™ and Standard Resolution modes [J].
Bosco, E ;
D'Agosta, L ;
Mancini, P ;
Traisci, G ;
D'Elia, C ;
Filipo, R .
ACTA OTO-LARYNGOLOGICA, 2005, 125 (02) :148-158
[4]   Short-term study of the effect of speech coding strategy on the auditory performance of pre- and post-lingually deafened adults implanted with the clarion CII® [J].
Filipo, R ;
Mancini, P ;
Ballantyne, D ;
Bosco, E ;
D'Elia, C .
ACTA OTO-LARYNGOLOGICA, 2004, 124 (04) :368-370
[5]  
FIRSZT JB, 2003, ADV BIONICS CORP
[6]   HiResolution™ and conventional sound processing in the HiResolution™ bionic ear:: Using appropriate outcome measures to assess speech recognition ability [J].
Koch, DB ;
Osberger, MJ ;
Segel, P ;
Kessler, D .
AUDIOLOGY AND NEURO-OTOLOGY, 2004, 9 (04) :214-223
[7]   DEVELOPMENT OF THE HEARING IN NOISE TEST FOR THE MEASUREMENT OF SPEECH RECEPTION THRESHOLDS IN QUIET AND IN NOISE [J].
NILSSON, M ;
SOLI, SD ;
SULLIVAN, JA .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1994, 95 (02) :1085-1099