Detection of a Meal Using Continuous Glucose Monitoring implications for an artificial β-cell

被引:87
作者
Dassau, Eyal [1 ]
Bequette, B. Wayne [2 ]
Buckingham, Bruce A. [3 ]
Doyle, Francis J., III [1 ]
机构
[1] Univ Calif Santa Barbara, Dept Chem Engn, Santa Barbara, CA 93106 USA
[2] Rensselaer Polytech Inst, Dept Chem & Biol Engn, Troy, NY USA
[3] Stanford Med Ctr, Stanford, CA USA
基金
美国国家卫生研究院;
关键词
SYSTEM; MODEL; CHILDREN; TIME;
D O I
10.2337/dc07-1293
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE - The purpose of this Study -,vas to introduce a novel meal detection algorithm (MDA) to be used as part of an artificial P-cell that uses a continuous glucose monitor (CGM). RESEARCH DESIGN AND METHODS - We developed our MDA on a dataset of 26 meal events using records from 19 children aged 1-6 years who Used the MiniMed CGMS Gold. We then applied this algorithm to CGM records from a DirecNet pilot study of the FreeStyle Navigator continuous glucose sensor. During a research center admission, breakfast insulin was withheld for 1 h, and discrete glucose levels were obtained every 10 min after the meal. RESULTS - Based on the Navigator readings, the MDA detected a meal at a mean time of 30 min from the onset of eating, at which time the mean serum glucose was 21 mg/dl above baseline (range 2-36 mg/dl), and >90% of meals were detected before the glucose had risen 40 mg/dl from baseline. CONCLUSIONS - The MDA will enable automated insulin dosing in response to meals, facilitating the development of an artificial pancreas.
引用
收藏
页码:295 / 300
页数:6
相关论文
共 21 条
[1]  
Bequette B Wayne, 2005, Diabetes Technol Ther, V7, P28, DOI 10.1089/dia.2005.7.28
[2]  
Bequette BW, 2004, P AMER CONTR CONF, P958
[3]   A system model of oral glucose absorption: Validation on gold standard data [J].
Dalla Man, Chiara ;
Camilleri, Michael ;
Cobelli, Claudio .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (12) :2472-2478
[4]  
Dassau E, 2007, DIABETES, V56, pA23
[5]   The medtronic MiniMed gold continuous glucose monitoring system: An effective means to discover hypo- and Hyperglycemia in children under 7 years of age [J].
Gandrud, Laura M. ;
Xing, Dongyuan ;
Kollman, Craig ;
Block, Jen M. ;
Kunselman, Betsy ;
Wilson, Darrell M. ;
Buckingham, Bruce A. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2007, 9 (04) :307-316
[6]   A new approach to diabetic control: Fuzzy logic and insulin pump technology [J].
Grant, Paul .
MEDICAL ENGINEERING & PHYSICS, 2007, 29 (07) :824-827
[7]   Continuous glucose monitoring and closed-loop systems [J].
Hovorka, R .
DIABETIC MEDICINE, 2006, 23 (01) :1-12
[8]   Roadmap to the artificial pancreas [J].
Hovorka, Roman ;
Wilinska, Malgorzata E. ;
Chassin, Ludovic J. ;
Dunger, David B. .
DIABETES RESEARCH AND CLINICAL PRACTICE, 2006, 74 :S178-S182
[9]  
Knobbe Edward J, 2005, Diabetes Technol Ther, V7, P15, DOI 10.1089/dia.2005.7.15
[10]  
KUUREKINSEY CR, 2006, P AM CONTR C MINN MN, P2957