Optimization of spoiled gradient-echo phase imaging for in vivo localization of a focused ultrasound beam

被引:160
作者
Chung, AH
Hynynen, K
Colucci, V
Oshio, K
Cline, HE
Jolesz, FA
机构
[1] HARVARD UNIV,MIT,DIV HLTH SCI & TECHNOL,CAMBRIDGE,MA 02139
[2] HARVARD UNIV,SCH MED,DEPT RADIOL,DIV MRI,BOSTON,MA
[3] GE CO,CTR CORP RES & DEV,SCHENECTADY,NY 12301
关键词
optimization; phase imaging; noninvasive thermometry; focused ultrasound surgery;
D O I
10.1002/mrm.1910360513
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The parameters of a spoiled gradient-echo (SPGR) pulse sequence have been optimized for in vivo localization of a focused ultrasound beam. Temperature elevation was measured by using the proton resonance frequency shift technique, and the phase difference signal-to-noise ratio (SNR Delta phi) was estimated in skeletal muscle and kidney cortex in 10 rabbits, Optimized parameters included the echo time equivalent to T-2* of the tissue, the longest repetition time possible with a 20-s sonication, and the flip angle equivalent to the Ernst angle. Optimal SPGR phase imaging can detect a sonication beam with a peak phase difference of 0.55 radian, which corresponds to a temperature elevation of 7.3 degrees C. The sonication beam can be localized within one voxel (0.6 x 0.6 x 5 mm(3)) at power levels that are below the threshold for thermal damage of the tissue.
引用
收藏
页码:745 / 752
页数:8
相关论文
共 32 条
  • [1] [Anonymous], 2012, ESSENTIALS MED ULTRA
  • [2] REAL-TIME MAGNETIC-RESONANCE-IMAGING OF LASER HEAT DEPOSITION IN TISSUE
    BLEIER, AR
    JOLESZ, FA
    COHEN, MS
    WEISSKOFF, RM
    DALCANTON, JJ
    HIGUCHI, N
    FEINBERG, DA
    ROSEN, BR
    MCKINSTRY, RC
    HUSHEK, SG
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1991, 21 (01) : 132 - 137
  • [3] CLINE H, 1992, J COMPUT ASSIST TOMO, V19, P956
  • [4] MR TEMPERATURE MAPPING OF FOCUSED ULTRASOUND SURGERY
    CLINE, HE
    HYNYNEN, K
    HARDY, CJ
    WATKINS, RD
    SCHENCK, JF
    JOLESZ, FA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (06) : 628 - 636
  • [5] ONLINE MONITORING OF ULTRASONIC SURGERY WITH MR-IMAGING
    DARKAZANLI, A
    HYNYNEN, K
    UNGER, EC
    SCHENCK, JF
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (03): : 509 - 514
  • [6] DEPOORTER J, 1995, MAGN RESON MED, V34, P359, DOI 10.1002/mrm.1910340313
  • [7] DEPOORTER J, 1995, MAGNET RESON MED, V33, P74, DOI 10.1002/mrm.1910330111
  • [8] DICKINSON RJ, 1986, J COMPUT ASSIST TOMO, V10, P468
  • [9] APPLICATION OF FOURIER TRANSFORM SPECTROSCOPY TO MAGNETIC RESONANCE
    ERNST, RR
    ANDERSON, WA
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1966, 37 (01) : 93 - +
  • [10] MR IMAGING OF THE LIVER, KIDNEYS, AND ADRENAL-GLANDS
    GLAZER, GM
    [J]. RADIOLOGY, 1988, 166 (02) : 303 - 312