Microbore high-performance liquid chromatographic method for measuring acetylcholine in microdialysis samples: Optimizing performance of platinum electrodes

被引:10
作者
Carter, AJ [1 ]
Kehr, J [1 ]
机构
[1] KAROLINSKA INST,DEPT NEUROSCI,S-17177 STOCKHOLM,SWEDEN
来源
JOURNAL OF CHROMATOGRAPHY B | 1997年 / 692卷 / 01期
关键词
platinum electrodes; acetylcholine;
D O I
10.1016/S0378-4347(96)00491-4
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A microbore high-performance liquid chromatographic method with electrochemical detection was applied to the measurement of acetylcholine in microdialysis samples. There was an excellent linear relationship (r = 0.99998) between the concentration of acetylcholine injected onto the column and the peak height (0.05-10 pmol/5 mu l). During the validation of this method, we noticed that the peak height for acetylcholine decreased over time, coupled with the appearance of a brown coating on the surface of the platinum electrode. Repeated measurement of acetylcholine standards which had been stored at 4 degrees C and -20 degrees C before and after cleaning the platinum electrode with ethanol or methanol indicated that the decrease in the peak height of acetylcholine is caused by a decrease in sensitivity of the electrode itself Results with a second microbore high-performance liquid chromatographic system confirmed these findings. On the basis of these results, we recommend that the platinum electrode is cleaned periodically with ethanol or methanol, and that quantitation is regularly calibrated with external acetylcholine.
引用
收藏
页码:207 / 212
页数:6
相关论文
共 15 条
[1]   MICROBORE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE MEASUREMENT OF DOPAMINE AND ITS METABOLITES - RECOMMENDATIONS FOR OPTIMAL SAMPLE COLLECTION AND STORAGE [J].
CARTER, AJ .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1994, 660 (01) :158-163
[2]  
CARTER AJ, 1995, J PHARMACOL EXP THER, V273, P637
[3]   A SIMPLE, SENSITIVE, AND ECONOMIC ASSAY FOR CHOLINE AND ACETYLCHOLINE USING HPLC, AND ENZYME REACTOR, AND AN ELECTROCHEMICAL DETECTOR [J].
DAMSMA, G ;
WESTERINK, BHC ;
HORN, AS .
JOURNAL OF NEUROCHEMISTRY, 1985, 45 (05) :1649-1652
[4]   MEASUREMENT OF ACETYLCHOLINE-RELEASE IN FREELY MOVING RATS BY MEANS OF AUTOMATED INTRACEREBRAL DIALYSIS [J].
DAMSMA, G ;
WESTERINK, BHC ;
DEVRIES, JB ;
VANDENBERG, CJ ;
HORN, AS .
JOURNAL OF NEUROCHEMISTRY, 1987, 48 (05) :1523-1528
[5]   CONTROLLED-POTENTIAL COULOMETRIC DETERMINATION OF HYDROGEN PEROXIDE [J].
HARRAR, JE .
ANALYTICAL CHEMISTRY, 1963, 35 (07) :893-&
[6]   Detection of basal acetylcholine release in the microdialysis of rat frontal cortex by high-performance liquid chromatography using a horseradish peroxidase-osmium redox polymer electrode with pre-enzyme reactor [J].
Kato, T ;
Liu, JK ;
Yamamoto, K ;
Osborne, PG ;
Niwa, O .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 1996, 682 (01) :162-166
[7]  
Lingane J.J., 1963, J ELECTROANAL CHEM, V5, P411, DOI 10.1016/0022-0728(63)80050-9
[8]  
Pettit HO, 1991, MICRODIALYSIS NEUROS, P117
[9]   ACETYLCHOLINE AND CHOLINE IN NEURONAL TISSUE MEASURED BY HPLC WITH ELECTROCHEMICAL DETECTION [J].
POTTER, PE ;
MEEK, JL ;
NEFF, NH .
JOURNAL OF NEUROCHEMISTRY, 1983, 41 (01) :188-194
[10]  
Taylor P, 1996, GOODMAN GILMANS PHAR, P161