Acquisition of electrophysiologic signals during magnetic resonance imaging

被引:10
作者
Parker, JM
Alger, JR
Woo, MA
Spriggs, D
Harper, RM
机构
[1] Univ Calif Los Angeles, Sch Med, Dept Neurobiol, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Dept Radiol Sci, Los Angeles, CA 90095 USA
[3] Univ Calif Los Angeles, Brain Res Inst, Los Angeles, CA 90095 USA
关键词
electroencephalogram; electrocardiogram; sleep; amplifier;
D O I
10.1093/sleep/22.8.1125
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
We describe a low cost system for acquiring electrophysiological signals during magnetic resonance imaging. The system consists of high common-mode-rejection and low noise operational amplifiers, coupled by fiber optic cables to a receiver located at the periphery of the magnetic field. The system minimizes noise introduction which would contaminate image signals.
引用
收藏
页码:1125 / 1126
页数:2
相关论文
共 1 条
  • [1] MONITORING THE PATIENTS EEG DURING ECHO-PLANAR MRI
    IVES, JR
    WARACH, S
    SCHMITT, F
    EDELMAN, RR
    SCHOMER, DL
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1993, 87 (06): : 417 - 420