Three-dimensional CT with a modified C-arm image intensifier: Feasibility

被引:88
作者
Linsenmaier, U
Rock, C
Euler, E
Wirth, S
Brandl, R
Kotsianos, D
Mutschler, W
Pfeifer, KJ
机构
[1] Univ Munich, Klinikum Univ Munchen Innenstadt, Dept Radiol, D-80336 Munich, Germany
[2] Univ Munich, Klinikum Univ Munchen Innenstadt, Dept Surg, D-80336 Munich, Germany
关键词
computed tomography (CT); technology; three-dimensional; extremities; CT;
D O I
10.1148/radiol.2241010646
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A portable C arm was modified for cone-beam computed tomography (CT). This three-dimensional (3D) CT imaging system facilitated the acquisition of fluoroscopic images during a 190degrees rotation and computed a 3D data cube (matrix, 256 X 256 X 256; scanning time, 100 seconds) with multiplanar image reformation. The high-contrast resolution, 0.9 line pairs per millimeter, was comparable; the low-contrast resolution, minimal; and the radiation dose, 600/6-80% lower, as compared with these parameters at spiral CT. The normal anatomy of small joints could be depicted, and the osteosynthesis screws in the talus were correctly identified. (C) RSNA, 2002.
引用
收藏
页码:286 / 292
页数:7
相关论文
共 21 条
[1]   Spiral CT vs incremental CT: Is spiral CT superior in imaging of the brain? [J].
Bahner, ML ;
Reith, W ;
Zuna, I ;
Engenhart-Cabillic, R ;
van Kaick, G .
EUROPEAN RADIOLOGY, 1998, 8 (03) :416-420
[2]  
BAUMHAUER JF, 1995, ORTHOP CLIN N AM, V26, P335
[3]   A CONE-BEAM RECONSTRUCTION ALGORITHM USING SHIFT-VARIANT FILTERING AND CONE-BEAM BACKPROJECTION [J].
DEFRISE, M ;
CLACK, R .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (01) :186-195
[4]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[5]  
Frodel J L Jr, 1999, Facial Plast Surg, V15, P83, DOI 10.1055/s-2008-1064303
[6]  
Grunert P, 1998, Comput Aided Surg, V3, P166, DOI 10.1002/(SICI)1097-0150(1998)3:4<166::AID-IGS6>3.3.CO
[7]  
2-5
[8]  
Haberland N, 2000, Comput Aided Surg, V5, P18, DOI 10.1002/(SICI)1097-0150(2000)5:1<18::AID-IGS3>3.3.CO
[9]  
2-K
[10]  
HAUSER H, 1999, EMERG RADIOL, V6, P61