Segmentation of cancellous bone from high-resolution computed tomography images: Influence on trablecular bone measurements

被引:28
作者
Elmoutaouakkil, A
Peyrin, F
Elkafi, J
Laval-Jeantet, AM
机构
[1] CNRS, UMR 5515, CREATIS, INSA, F-69621 Villeurbanne, France
[2] ESRF, F-38043 Grenoble, France
[3] Univ Chouaib Doukali, El Jadida, Morocco
关键词
high-resolution computed tomography; quantification; segmentation; tarbecular bone structure;
D O I
10.1109/TMI.2002.1000259
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The quantification of cancellous bone network from computed tomography (CT) images requires a segmentation step which is crucial and difficult because of the partial volume effect in CT images. In this paper, we present and evaluate a new approach for segmenting cancellous bone network from high-resolution CT (HRCT) slices. The idea is first to detect a skeleton from the crest lines of the structure and then to thicken it to extract the whole bone structure by satisfying local neighborhood constraints. The segmentation requires the adjustment of relative and not absolute parameters like most methods. We quantified the influence of these parameters on architectural measurements. Results were first validated by using a physical phantom and then examined on a series of 12 HRCT images of human lumbar vertebra of different ages. We demonstrated that the choice of segmentation parameters yielded important variability on architectural measurements (up to 20%), but less variability than a more commonly used approach. This stresses the importance of settle on the segmentation parameters for once, which is possible with the proposed method.
引用
收藏
页码:354 / 362
页数:9
相关论文
共 26 条
[1]   MEASUREMENT OF ANISOTROPIC VERTEBRAL TRABECULAR BONE LOSS DURING AGING BY QUANTITATIVE IMAGE-ANALYSIS [J].
BERGOT, C ;
LAVALJEANTET, AM ;
PRETEUX, F ;
MEUNIER, A .
CALCIFIED TISSUE INTERNATIONAL, 1988, 43 (03) :143-149
[2]  
BEUCHER S, 1978, LECT NOTES BIOMATH, V23, P311
[3]  
BOIVIN B, 1992, HISTOMORPHOMETRIC ME, P137
[4]   CT IMAGE-ANALYSIS OF THE VERTEBRAL TRABECULAR NETWORK INVIVO [J].
CHEVALIER, F ;
LAVALJEANTET, AM ;
LAVALJEANTET, M ;
BERGOT, C .
CALCIFIED TISSUE INTERNATIONAL, 1992, 51 (01) :8-13
[5]  
COEQUEREZ JP, 1995, ANAL IMAGES FILTRAGE
[6]   CANCELLOUS BONE-STRUCTURE - ANALYSIS OF HIGH-RESOLUTION CT IMAGES WITH THE RUN-LENGTH METHOD [J].
DURAND, EP ;
RUEGSEGGER, P .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (01) :133-139
[7]   HIGH SPATIAL-RESOLUTION IMAGING OF BONE-MINERAL USING COMPUTED MICROTOMOGRAPHY - COMPARISON WITH MICRORADIOGRAPHY AND UNDECALCIFIED HISTOLOGIC SECTIONS [J].
ENGELKE, K ;
GRAEFF, W ;
MEISS, L ;
HAHN, M ;
DELLING, G .
INVESTIGATIVE RADIOLOGY, 1993, 28 (04) :341-349
[8]   THE DIRECT EXAMINATION OF 3-DIMENSIONAL BONE ARCHITECTURE INVITRO BY COMPUTED-TOMOGRAPHY [J].
FELDKAMP, LA ;
GOLDSTEIN, SA ;
PARFITT, AM ;
JESION, G ;
KLEEREKOPER, M .
JOURNAL OF BONE AND MINERAL RESEARCH, 1989, 4 (01) :3-11