MD-Logic Overnight Control for 6 Weeks of Home Use in Patients With Type 1 Diabetes: Randomized Crossover Trial

被引:130
作者
Nimri, Revital [1 ]
Muller, Ido [1 ]
Atlas, Eran [1 ]
Miller, Shahar [1 ]
Fogel, Aviel [1 ]
Bratina, Natasa [2 ]
Kordonouri, Olga [3 ]
Battelino, Tadej [2 ,4 ]
Danne, Thomas [3 ]
Phillip, Moshe [1 ,5 ]
机构
[1] Schneider Childrens Med Ctr Israel, Natl Ctr Childhood Diabet, Jesse Z & Sara Lea Shafer Inst Endocrinol & Diabe, Petah Tiqwa, Israel
[2] Univ Childrens Hosp, Univ Med Ctr, Dept Pediat Endocrinol Diabet & Metab, Ljubljana, Slovenia
[3] BULT Kinder & Jugendkrankenhaus, Diabet Ctr Children & Adolescents, Hannover, Germany
[4] Univ Ljubljana, Fac Med, Ljubljana, Slovenia
[5] Tel Aviv Univ, Sackler Fac Med, IL-69978 Tel Aviv, Israel
关键词
AUGMENTED PUMP THERAPY; LOOP INSULIN DELIVERY; ARTIFICIAL PANCREAS; GLUCOSE CONTROL; ADULTS; CHILDREN; SYSTEM; HYPOGLYCEMIA; FEASIBILITY; ADOLESCENTS;
D O I
10.2337/dc14-0835
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
OBJECTIVE We evaluated the effect of the MD-Logic system on overnight glycemic control at patients' homes. RESEARCH DESIGN AND METHODS Twenty-four patients (aged 12-43 years; average A(1c) 7.5 +/- 0.8%, 58.1 +/- 8.4 mmol/mol) were randomly assigned to participate in two overnight crossover periods, each including 6 weeks of consecutive nights: one under closed loop and the second under sensor-augmented pump (SAP) therapy at patients' homes in real-life conditions. The primary end point was time spent with sensor glucose levels below 70 mg/dL (3.9 mmol/L) overnight. RESULTS Closed-loop nights significantly reduced time spent in hypoglycemia (P = 0.02) and increased the percentage of time spent in the target range of 70-140 mg/dL (P = 0.003) compared with nights when the SAP therapy was used. The time spent in substantial hyperglycemia above 240 mg/dL was reduced by a median of 52.2% (interquartile range [IQR] 4.8, 72.9%; P = 0.001) under closed-loop control compared with SAP therapy. Overnight total insulin doses were lower in the closed-loop nights compared with the SAP nights (P = 0.04). The average daytime glucose levels after closed-loop operation were reduced by a median of 10.0 mg/dL (IQR - 2.7, 19.2; P = 0.017) while lower total insulin doses were used (P = 0.038). No severe adverse events occurred during closed-loop control; there was a single event of severe hypoglycemia during a control night. CONCLUSIONS The long-term home use of automated overnight insulin delivery by the MD-Logic system was found to be a feasible, safe, and an effective tool to reduce nocturnal hypoglycemia and improve overnight glycemic control in subjects with type 1 diabetes.
引用
收藏
页码:3025 / 3032
页数:8
相关论文
共 28 条
[1]
MD-Logic Artificial Pancreas System A pilot study in adults with type 1 diabetes [J].
Atlas, Eran ;
Nimri, Revital ;
Miller, Shahar ;
Grunberg, Eli A. ;
Phillip, Moshe .
DIABETES CARE, 2010, 33 (05) :1072-1076
[2]
The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial [J].
Battelino, T. ;
Conget, I. ;
Olsen, B. ;
Schuetz-Fuhrmann, I. ;
Hommel, E. ;
Hoogma, R. ;
Schierloh, U. ;
Sulli, N. ;
Bolinder, J. .
DIABETOLOGIA, 2012, 55 (12) :3155-3162
[3]
Effect of Continuous Glucose Monitoring on Hypoglycemia in Type 1 Diabetes [J].
Battelino, Tadej ;
Phillip, Moshe ;
Bratina, Natasa ;
Nimri, Revital ;
Oskarsson, Per ;
Bolinder, Jan .
DIABETES CARE, 2011, 34 (04) :795-800
[4]
The T1D Exchange Clinic Registry [J].
Beck, Roy W. ;
Tamborlane, William V. ;
Bergenstal, Richard M. ;
Miller, Kellee M. ;
DuBose, Stephanie N. ;
Hall, Callyn A. .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2012, 97 (12) :4383-4389
[6]
Threshold-Based Insulin-Pump Interruption for Reduction of Hypoglycemia [J].
Bergenstal, Richard M. ;
Klonoff, David C. ;
Garg, Satish K. ;
Bode, Bruce W. ;
Meredith, Melissa ;
Slover, Robert H. ;
Ahmann, Andrew J. ;
Welsh, John B. ;
Lee, Scott W. ;
Kaufman, Francine R. .
NEW ENGLAND JOURNAL OF MEDICINE, 2013, 369 (03) :224-232
[7]
Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes [J].
Bergenstal, Richard M. ;
Tamborlane, William V. ;
Ahmann, Andrew ;
Buse, John B. ;
Dailey, George ;
Davis, Stephen N. ;
Joyce, Carol ;
Peoples, Tim ;
Perkins, Bruce A. ;
Welsh, John B. ;
Willi, Steven M. ;
Wood, Michael A. .
NEW ENGLAND JOURNAL OF MEDICINE, 2010, 363 (04) :311-320
[8]
The Effects of Inpatient Hybrid Closed-Loop Therapy Initiated Within 1 Week of Type 1 Diabetes Diagnosis [J].
Buckingham, Bruce A. ;
Beck, Roy W. ;
Ruedy, Katrina J. ;
Cheng, Peiyao ;
Kollman, Craig ;
Weinzimer, Stuart A. ;
DiMeglio, Linda A. ;
Bremer, Andrew A. ;
Slover, Robert ;
Cantwell, Martin ;
Tsalikian, Eva ;
Tansey, Michael J. ;
Coffey, Julie ;
Cabbage, Joanne ;
Salamati, Sara ;
Mauras, Nelly ;
Fox, Larry A. ;
Englert, Kim ;
Permuy, Joe ;
Sikes, Kaitlin ;
Buckingham, Bruce A. ;
Wilson, Darrell M. ;
Clinton, Paula ;
Caswell, Kimberly ;
Weinzimer, Stuart A. ;
Tamborlane, William V. ;
Sherr, Jennifer ;
Steffen, Amy ;
Weyman, Kate ;
Zgorski, Melinda ;
Tichy, Eileen ;
White, Neil H. ;
Arbelaez, Ana Maria ;
Levandoski, Lucy ;
Starnes, Angie ;
Skyler, Jay S. ;
Greenbaum, Carla J. ;
Kenyon, Norma S. ;
Rafkin, Lisa ;
Santiago, Irene ;
Sosenko, Jay M. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (05) :401-408
[9]
Closed-Loop Insulin Therapy Improves Glycemic Control in Children Aged <7 Years A randomized controlled trial [J].
Dauber, Andrew ;
Corcia, Liat ;
Safer, Jason ;
Agus, Michael S. D. ;
Einis, Sara ;
Steil, Garry M. .
DIABETES CARE, 2013, 36 (02) :222-227
[10]
Closed-Loop Basal Insulin Delivery Over 36 Hours in Adolescents With Type 1 Diabetes [J].
Elleri, Daniela ;
Allen, Janet M. ;
Kumareswaran, Kavita ;
Leelarathna, Lalantha ;
Nodale, Marianna ;
Caldwell, Karen ;
Cheng, Peiyao ;
Kollman, Craig ;
Haidar, Ahmad ;
Murphy, Helen R. ;
Wilinska, Malgorzata E. ;
Acerini, Carlo L. ;
Dunger, David B. ;
Hovorka, Roman .
DIABETES CARE, 2013, 36 (04) :838-844