Empirical procedure for the reduction of mixed-matrix effects in inductively coupled plasma atomic emission spectrometry using an internal standard and proportional correction

被引:36
作者
deBoer, JLM [1 ]
Velterop, M [1 ]
机构
[1] UNIV UTRECHT,DEPT GEOCHEM,NL-3508 TA UTRECHT,NETHERLANDS
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1996年 / 356卷 / 06期
关键词
D O I
10.1007/s0021663560362
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The matrix effects for single-matrix component (Na or Ca) and mixed-matrix component (Na plus Ca) systems were investigated for an inductively coupled plasma with radial viewing and employing ultrasonic nebulization. Correction for matrix effects was carried out using the internal standard ratio method and a method named 'proportional correction', both with scandium and yttrium as internal standard elements. The best correction was observed using proportional correction both for the single-matrix component systems and for the mixed-matrix component system.
引用
收藏
页码:362 / 370
页数:9
相关论文
共 33 条
[31]   THE EFFECT OF EASILY IONIZED ELEMENTS ON ANALYTE EMISSION EFFICIENCY IN INDUCTIVELY-COUPLED PLASMA SPECTROMETRY [J].
WU, M ;
HIEFTJE, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1994, 49 (02) :149-161
[32]   INTERFERENCE BY MINERAL ACIDS IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
YOSHIMURA, E ;
SUZUKI, H ;
YAMAZAKI, S ;
TODA, S .
ANALYST, 1990, 115 (02) :167-171
[33]  
Zaidel' A.N., 1970, TABLES SPECTRAL LINE