Doppler sensors and harnesses for cardiac and peripheral arterial flow monitoring

被引:12
作者
Arbeille, P
机构
[1] Unité Médecine and Physiologie Spatiales, CHU Trousseau, Tours
[2] Médecine Nucléaire and Ultrasons (UMPS), CHU Trousseau
关键词
Doppler; skin sensor; harness; cardiovascular; monitoring;
D O I
10.1016/S0301-5629(96)00225-6
中图分类号
O42 [声学];
学科分类号
070206 [声学]; 082403 [水声工程];
摘要
The objective of the present work was to design Doppler sensors and harnesses for monitoring in real time the cardiac stroke volume and output, the cerebral flow volume and resistance and the lower limb arterial flow and resistance changes. For the middle cerebral artery investigation we used a 2-MHz transcranial pulsed wave (PW) Doppler probe (commercial probe) mounted on a rotula, fixed on the horizontal branch of a headset designed like an audio headset. The arch of the headset passed over the top of the skull. For the common carotid investigation, a 4-MHz continuous-wave (CW) or PW flat Doppler probe was inserted in a circular silicone support of 5-cm diameter and 0.6-cm thickness. This soft support could adapt to the irregular surface of the neck, and the silicone material, which is fairly adhesive to the skin, made the system stable. The transducers were preoriented at 45 degrees from the support, which provided an acceptable orientation of the Doppler beam. The aortic Doppler harness consisted of a 2-MHz PW Doppler probe, mounted on a rotula fixed on a plastic rigid support 7 X 1.5 cm(2) in area (parallel to the sternum) whose length could be changed (4-7 cm) to localize the rotula and its sensor on the suprasternal area. This flat segment was fixed on a rotating platform, part of a solid square plastic support (10 X 10 cm(2)) placed on the upper part of the sternum. This system was maintained by elastic bands passing around the shoulder and the chest. The femoral Doppler harness consisted of a 4-MHz CW or PW Bat Doppler probe, inserted in a flat and rectangular rigid plastic support 10 X 3 cm(2) in area. This rigid support was well adapted to the flat surface of the internal part of the thigh, which made the system stable. Two elastic bands passed around the thigh and the abdomen, which avoided any translation of the support and kept the sensor in contact with the skin. The transducers were preoriented at 45 degrees from the support, which provided an acceptable orientation of the Doppler beam. The whole Doppler monitoring system was successfully tested during orthostatic tests,such as lower body negative pressure or tilt table tests performed in pre and post head-down tilt experiments and space flights, with the objective of assessing and quantifying the cardiac and peripheral arterial hemodynamic changes associated with orthostatic intolerance. (C) 1997 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:415 / 423
页数:9
相关论文
共 12 条
[1]
NON-INVASIVE TRANSCRANIAL DOPPLER ULTRASOUND RECORDING OF FLOW VELOCITY IN BASAL CEREBRAL-ARTERIES [J].
AASLID, R ;
MARKWALDER, TM ;
NORNES, H .
JOURNAL OF NEUROSURGERY, 1982, 57 (06) :769-774
[2]
ADAMSON SL, 1990, ULTRASOUND MED BIOL, V6, P19
[3]
NEW IMPLANTED DOPPLER SENSORS FOR THE ASSESSMENT OF THE MAIN FETAL HEMODYNAMICS [J].
ARBEILLE, P ;
BERSON, M ;
MAULIK, D ;
LOCATELLI, A ;
BODARD, S .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1992, 18 (01) :97-103
[4]
ARBEILLE P, 1995, AVIAT SPACE ENVIR MD, V66, P131
[5]
VASCULAR-RESISTANCE QUANTIFICATION IN HIGH-FLOW RESISTANCE AREAS USING THE DOPPLER METHOD [J].
ARBEILLE, P ;
BERSON, M ;
ACHAIBOU, F ;
BODARD, S ;
LOCATELLI, A .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (03) :321-328
[6]
ARBEILLE P, 1996, ACTA ASTRONAUT, V36, P753
[7]
ARBEILLE P, 1995, ULTRASOUND MED BIOL, V21, P101
[8]
ARBEILLE P, 1996, IN PRESS J GRAVITATI
[9]
COATS AJS, 1990, JANSSEN BIOMEDICAL S, P49
[10]
LETRAON AP, 1995, AVIAT SPACE ENVIR MD, V66, P982