A retrospective case study to assess the value of the implantable loop recorder for the investigation of undiagnosed syncope

被引:15
作者
Ashby, DT
Cehic, DA
Disney, PJS
Mahar, LJ
Young, GD
机构
[1] Royal Adelaide Hosp, Cardiovasc Invest Unit, Adelaide, SA 5000, Australia
[2] Adelaide Cardiol, Adelaide, SA, Australia
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2002年 / 25卷 / 08期
关键词
D O I
10.1046/j.1460-9592.2002.01200.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
If not diagnosed by history, examination, or ECG, the diagnosis of syncope can be difficult with a low yield from echocardiography, ambulatory ECG recording, electrophysiological study, and tilt table testing. During 2 years, 48 patients with unexplained syncope or presyncope from three hospitals in one city underwent the implantation of a Medtronic Reveal implantable loop recorder capable of cardiac monitoring for 14 months. All patients had at least two prior episodes of syncope or presyncope. Fifty-two percent of patients had electrophysiological studies, all of which were negative. The implantable loop recorder remained implanted until a diagnostic event was recorded, or until the end of the battery life. After a mean follow-up of 5.6 +/- 5.7 months, symptoms reoccurred in 25 (52.1%) patients at a mean of 2.8 +/- 2.1 months after insertion of an implantable loop recorder. No further symptoms occurred in 23 (47.9%) patients. Of the 25 patients who had a symptom and recorded an event, an arrhythmia was seen in 10 (40%) patients. Seven patients had bradycardia; 4 with profound sinus bradycardia/sinus arrest, 1 with complete heart block, and 2 in association with the cardioinhibitory component of vasovagal syncope. Three patients had tachycardias; two with supraventricular tachycardia and one with atrial flutter. Fifteen (60%) of the 25 patients who activated their device due to syncope or presyncope were in sinus rhythm during the event. The implantable loop recorder was effective in making a cardiological or noncardiological diagnosis for unexplained syncope or presyncope in 52.1% of the patients.
引用
收藏
页码:1200 / 1205
页数:6
相关论文
共 19 条
[1]   LONG-TERM PROGNOSIS OF PATIENTS UNDERGOING ELECTROPHYSIOLOGIC STUDIES FOR SYNCOPE OF UNKNOWN ORIGIN [J].
BASS, EB ;
ELSON, JJ ;
FOGOROS, RN ;
PETERSON, J ;
ARENA, VC ;
KAPOOR, WN .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 62 (17) :1186-1191
[2]   ARRHYTHMIAS DETECTED BY AMBULATORY MONITORING - LACK OF CORRELATION WITH SYMPTOMS OF DIZZINESS AND SYNCOPE [J].
CLARK, PI ;
GLASSER, SP ;
SPOTO, E .
CHEST, 1980, 77 (06) :722-725
[3]   ADENOSINE FOR PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA - DOSE RANGING AND COMPARISON WITH VERAPAMIL - ASSESSMENT IN PLACEBO-CONTROLLED, MULTICENTER TRIALS [J].
DIMARCO, JP ;
MILES, W ;
AKHTAR, M ;
MILSTEIN, S ;
SHARMA, AD ;
PLATIA, E ;
MCGOVERN, B ;
SCHEINMAN, MM ;
GOVIER, WC .
ANNALS OF INTERNAL MEDICINE, 1990, 113 (02) :104-110
[4]   DIAGNOSIS AND NATURAL-HISTORY OF SYNCOPE AND THE ROLE OF INVASIVE ELECTROPHYSIOLOGIC TESTING [J].
KAPOOR, WN ;
HAMMILL, SC ;
GERSH, BJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 63 (11) :730-734
[5]  
KAPOOR WN, 1991, AM J MED, V90, P91
[6]   A PROSPECTIVE EVALUATION AND FOLLOW-UP OF PATIENTS WITH SYNCOPE [J].
KAPOOR, WN ;
KARPF, M ;
WIEAND, S ;
PETERSON, JR ;
LEVEY, GS .
NEW ENGLAND JOURNAL OF MEDICINE, 1983, 309 (04) :197-204
[7]   The high cost of syncope: Cost implications of a new insertable loop recorder in the investigation of recurrent syncope [J].
Krahn, AD ;
Klein, GJ ;
Yee, R ;
Manda, V .
AMERICAN HEART JOURNAL, 1999, 137 (05) :870-877
[8]   Use of an extended monitoring strategy in patients with problematic syncope [J].
Krahn, AD ;
Klein, GJ ;
Yee, R ;
Takle-Newhouse, T ;
Norris, C .
CIRCULATION, 1999, 99 (03) :406-410
[9]   THE ETIOLOGY OF SYNCOPE IN PATIENTS WITH NEGATIVE TILT-TABLE AND ELECTROPHYSIOLOGICAL TESTING [J].
KRAHN, AD ;
KLEIN, GJ ;
NORRIS, C ;
YEE, R .
CIRCULATION, 1995, 92 (07) :1819-1824
[10]   Randomized assessment of syncope trial - Conventional diagnostic testing versus a prolonged monitoring strategy [J].
Krahn, AD ;
Klein, GJ ;
Yee, R ;
Skanes, AC .
CIRCULATION, 2001, 104 (01) :46-51