Comparison of lateral locking plate and antiglide plate for fixation of distal fibular fractures in osteoporotic bone: A biomechanical study

被引:129
作者
Minihane, Keith P.
Lee, Chris
Ahn, Chulhyun
Zhang, Li-Qun
Merk, Bradley R.
机构
[1] Northwestern Univ, Sch Med, Dept Orthopaed Surg, Chicago, IL 60611 USA
[2] Rehabil Inst Chicago, Chicago, IL 60611 USA
关键词
antiglide; locking plate; fibula fracture; osteoporosis; ankle fracture; biomechanical;
D O I
10.1097/01.bot.0000245684.96775.82
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: The purpose of this study was to compare the biomechanical properties of posterolateral antiglide plating and lateral locked plating for fixation of displaced short oblique fractures of the fibula in osteoporotic bone. Methods: Short oblique fractures of the distal fibula at the level of the syndesmosis were simulated with a fibular osteotomy and ligamentous sectioning in 18 paired fresh frozen ankles. The fractures were fixed with either a lateral locking plate with an independent lag screw or a posterolateral antiglide plate with a lag screw through the plate. The specimens were tested under a torsional load to failure. The torque to failure, angular rotation at failure, and construct stiffness of the two groups were compared. Results: The torque to failure and construct stiffness were significantly greater on the side with the posterolateral antiglide plate than on the side with the the lateral locking plate (P = 0.01 and 0.005, respectively). Conclusions: The posterolateral antiglide plate demonstrated improved biomechanical stability as compared to the lateral locking plate in osteoporotic bone. In situations where fixation needs to be optimized, use of an antiglide plate may be favored over a lateral locking plate construct.
引用
收藏
页码:562 / 566
页数:5
相关论文
共 31 条
[1]
Ali M S, 1987, J Orthop Trauma, V1, P275, DOI 10.1097/00005131-198701040-00002
[2]
[Anonymous], 1982, SPECIAL TECHNIQUES I
[3]
INTRAMEDULLARY SCREW FIXATION OF LATERAL MALLEOLUS FRACTURES [J].
BANKSTON, AB ;
ANDERSON, LD ;
NIMITYONGSKUL, P .
FOOT & ANKLE INTERNATIONAL, 1994, 15 (11) :599-607
[4]
The benefit of aggressive lipid lowering [J].
Brown, WV .
ATHEROSCLEROSIS SUPPLEMENTS, 2000, 1 (01) :15-19
[5]
An augmented fixation method for distal fibular fractures in elderly patients: A biomechanical evaluation [J].
Dunn, WR ;
Easley, ME ;
Parks, BG ;
Trnka, HJ ;
Schon, LC .
FOOT & ANKLE INTERNATIONAL, 2004, 25 (03) :128-131
[6]
Biomechanics of locked plates and screws [J].
Egol, KA ;
Kubiak, EN ;
Fulkerson, E ;
Kummer, FJ ;
Koval, KJ .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) :488-493
[7]
Gardner Michael J, 2004, Am J Orthop (Belle Mead NJ), V33, P439
[8]
Innovations in locking plate technology [J].
Haidukewych, GJ .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2004, 12 (04) :205-212
[9]
Higgins JPT, 2011, CLIN ORTHOP RELAT R
[10]
A biomechanical evaluation of methods of distal humerus fracture fixation using locking compression plates versus conventional reconstruction plates [J].
Korner, J ;
Diederichs, G ;
Arzdorf, M ;
Lill, H ;
Josten, C ;
Schneider, E ;
Linke, B .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (05) :286-293