Between March 1990 and May 1991 we performed 85 primary total hip replacements in 74 patients using the Landos Atoll hydroxyapatite (HA)-coated cup and the Corail HA-coated stem. The patients were followed up for a mean of ten years. Of the 85 cups, 26 (31%) have already been revised and a further six are radiologically unstable and awaiting revision. Two femoral stems have been revised for infection without loosening. The retrieved acetabular cups were studied by SEM and image-processing techniques to quantify the amount of residual HA on the cup. This was correlated with the clinical variables and modes of failure. The residual HA (as a percentage of the surface) on the loose cups correlated negatively with the duration of implantation (r = -0.732, p < 0.001). Six cups were stable at revision and had more residual HA coating than those which were loose (p < 0.01). The rate of failure of the Landos Atoll HA-coated, smooth hemispherical cup with screw fixation is unacceptably high. Resorption of the HA coating is markedly increased in loose cups compared with stable cups. HA coating cannot substitute for stable mechanical fixation.
引用
收藏
页码:641 / 646
页数:6
相关论文
共 22 条
[1]
HYDROXYAPATITE-COATED FEMORAL STEMS - HISTOLOGICAL ANALYSIS OF COMPONENTS RETRIEVED AT AUTOPSY [J].
BAUER, TW ;
GEESINK, RCT ;
ZIMMERMAN, R ;
MCMAHON, JT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (10) :1439-1452
[2]
BAUER TW, 1993, J ARTHROPLASTY, V8, P167, DOI 10.1016/S0883-5403(09)80010-7
[3]
Hydroxyapatite in total hip arthroplasty -: Clinical results and critical issues [J].
Capello, WN ;
D 'Antonio, JA ;
Manley, MT ;
Feinberg, JR .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (355) :200-211
[4]
HYDROXYLAPATITE COATING OF POROUS IMPLANTS IMPROVES BONE INGROWTH AND INTERFACE ATTACHMENT STRENGTH [J].
COOK, SD ;
THOMAS, KA ;
DALTON, JE ;
VOLKMAN, TK ;
WHITECLOUD, TS ;
KAY, JF .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (08) :989-1001
[5]
THE EFFECT OF OPERATIVE FIT AND HYDROXYAPATITE COATING ON THE MECHANICAL AND BIOLOGICAL RESPONSE TO POROUS IMPLANTS [J].
DALTON, JE ;
COOK, SD ;
THOMAS, KA ;
KAY, JF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (01) :97-110
[6]
MECHANICAL AND HISTOLOGICAL-EVALUATION OF HYDROXYAPATITE-COATED, TITANIUM-COATED AND GRIT-BLASTED SURFACES UNDER WEIGHT-BEARING CONDITIONS [J].
DAVID, A ;
EITENMULLER, J ;
MUHR, G ;
POMMER, A ;
BAR, HF ;
OSTERMANN, PAW ;
SCHILDHAUER, TA .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1995, 114 (02) :112-118
[7]
BONDING OF BONE TO APATITE-COATED IMPLANTS [J].
GEESINK, RGT ;
DEGROOT, K ;
KLEIN, CPAT .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (01) :17-22
[8]
Total hip arthroplasty with hydroxyapatite-coated prostheses [J].
Jaffe, WL ;
Scott, DF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (12) :1918-1934
[9]
HISTOLOGY OF TISSUE ADJACENT TO AN HAC-COATED FEMORAL PROSTHESIS - A CASE-REPORT [J].
LINTNER, F ;
BOHM, G ;
HUBER, M ;
SCHOLZ, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (05) :824-830
[10]
The results at nine to twelve years of the use of a hydroxyapatite-coated femoral stem [J].
McNally, SA ;
Shepperd, JAN ;
Mann, CV ;
Walczak, JP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2000, 82B (03) :378-382